利用信号平均的压缩传感技术提高氟-19 MRI的灵敏度并减少运动干扰

Compressed Sensing with Signal Averaging for Improved Sensitivity and Motion Artifact Reduction in Fluorine-19 MRI

Emeline Darçot, Jérôme Yerly, Tom Hilbert, Roberto Colotti, Elena Najdenovska, Tobias Kober, Matthias Stuber, Ruud B. van Heeswijk

摘要:

注射的全氟化碳乳剂(PFC)的氟-19(19 F)MRI允许炎症和细胞追踪的非侵入性定量,但遭受低信噪比和延长的扫描时间。为了解决这一局限性,我们测试了一个假设,即与非平均全采样数据集相比,将特定欠采样机制与信号平均相结合的19 F MRI脉冲序列具有更高的灵敏度和对运动伪影的鲁棒性,当两者都用压缩感知重建时。为此,进行了数值模拟和体模实验,以表征欠采样模式的点扩散函数(PSF)和具有成对数量的x信号平均值和加速因子x(NAx-AFx)的采集重建策略对噪声的脆弱性。在所有研究的噪声水平下,采集重建策略的DSC强烈依赖于正则化参数和加速因子。在幻影,运动鲁棒性的NA 8-AF 8欠采样模式与NA 1-AF 1进行了评估与模拟和真实的运动。用Dice相似系数(DSC)评估差异,对于模拟和真实循环运动,与NA 1-AF 1策略相比,NA 8-AF 8策略的差异始终更高(P<0.001)。在注射全氟聚醚的小鼠(n=2)中体内验证了两种采集重建策略。与NA 1-AF 1策略相比,这些图像显示了NA 8-AF 8策略的肝脏轮廓更清晰。总之,我们验证了以下假设:在19 F MRI中,与非平均全采样数据集相比,欠采样和平均的组合提高了灵敏度和对运动伪影的鲁棒性,当两者都用压缩感知重建时。

英文摘要:

Fluorine-19 (19F) MRI of injected perfluorocarbon emulsions (PFCs) allows for the non-invasive quantification of inflammation and cell tracking, but suffers from a low signal-to-noise ratio and extended scan time. To address this limitation, we tested the hypothesis that a 19F MRI pulse sequence that combines a specific undersampling regime with signal averaging has increased sensitivity and robustness against motion artifacts compared to a non-averaged fully-sampled dataset, when both are reconstructed with compressed sensing. To this end, numerical simulations and phantom experiments were performed to characterize the point spread function (PSF) of undersampling patterns and the vulnerability to noise of acquisition-reconstruction strategies with paired numbers of x signal averages and acceleration factor x (NAx-AFx). At all investigated noise levels, the DSC of the acquisition-reconstruction strategies strongly depended on the regularization parameters and acceleration factor. In phantoms, motion robustness of an NA8-AF8 undersampling pattern versus NA1-AF1 was evaluated with simulated and real motions. Differences were assessed with Dice similarity coefficients (DSC), and were consistently higher for NA8-AF8 compared to NA1-AF1 strategy, for both simulated and real cyclic motions (P<0.001). Both acquisition-reconstruction strategies were validated in vivo in mice (n=2) injected with perfluoropolyether. These images displayed a sharper delineation of the liver with the NA8-AF8 strategy than with the NA1-AF1 strategy. In conclusion, we validated the hypothesis that in 19F MRI, the combination of undersampling and averaging improves both the sensitivity and the robustness against motion artifacts compared to a non-averaged fully-sampled dataset, when both are reconstructed with compressed sensing.

机构、总结、中文摘要均由AI分析而来,可能存在错误,请以论文正文为准。

↑