在跳动心脏模型中结合生理运动的TAVR支架设计计算机疲劳优化

In Silico Fatigue Optimization of TAVR Stent Designs with Physiological Motion in a Beating Heart Model

Kyle Baylous, Ryan Helbock, Brandon Kovarovic, Salwa Anam, Marvin Slepian, Danny Bluestein

AI总结:

本研究提出一种基于跳动心脏模型的TAVR支架计算机疲劳分析方法,通过有限元分析评估多种支架在动态载荷下的抗疲劳性,为器械耐久性评估提供了经济高效的体外试验替代方案。

摘要:

TAVR向年轻、低风险患者的快速扩展引发了对器械耐久性的担忧。必然地,延长的支架寿命对于新一代器械将变得更为关键。通常用于TAVR支架疲劳测试的体外方法排除了跳动心脏的影响。我们提出了一种更真实的计算机支架疲劳分析,利用跳动心脏模型,其中TAVR支架承受复杂的、非均匀的动态载荷。使用自膨胀式镍钛合金26-mm CoreValve和Evolut R器械的支架模型,以及27-mm PolyV-2,在跳动的SIMULIA Living Heart Human Model中模拟了虚拟TAVR部署。监测了三个心动周期的支架变形,并通过ABAQUS/Explicit的有限元分析评估了镍钛合金支架的抗疲劳性。在所有模型中,均存在应变超过疲劳失效的单元。PolyV-2支架的失效单元要少得多,因为其支柱经过优化以减少支架关节处的应变,在支架冠部和腰部单元中实现了更好的抗疲劳性。由于载荷条件不同,不同支架部分表现出显著不同的抗疲劳性。本研究展示了在跳动心脏内部署器械的先进计算机分析的实用性,其模拟了体内载荷,为人类或动物试验提供了经济高效的替代方案,并建立了一个评估器械设计对器械耐久性影响的平台。此处指出的TAVR支架有限的疲劳寿命凸显了一种临床并发症,随着更年轻、更低风险的TAVR患者年龄增长,这种并发症最终可能出现。

英文摘要:

The rapid expansion of TAVR to younger, low-risk patients raises concerns regarding device durability. Necessarily, extended stent lifetime will become more critical for new generation devices. In vitro methods commonly used for TAVR stent fatigue testing exclude the effects of the beating heart. We present a more realistic in silico stent fatigue analysis utilizing a beating heart model in which TAVR stents experience complex, nonuniform dynamic loading. Virtual TAVR deployments were simulated in the SIMULIA Living Heart Human Model of a beating heart using stent models of the self-expandable nitinol 26-mm CoreValve and Evolut R devices, and a 27-mm PolyV-2. Stent deformation was monitored over three cardiac cycles, and fatigue resistance was evaluated for the nitinol stents using finite element analysis via ABAQUS/Explicit. In all models, there were elements in which strains exceeded fatigue failure. The PolyV-2 stent had far fewer failing elements since its struts were optimized to reduce the strain in stent joints, achieving better fatigue resistance in the stent crown and waist elements. Different stent sections showed markedly different fatigue resistance due to the varying loading conditions. This study demonstrates the utility of advanced in silico analysis of devices deployed within a beating heart that mimics in vivo loading, offering a cost-effective alternative to human or animal trials and establishing a platform to assess the impact of device design on device durability. The limited fatigue life of TAVR stents indicated here highlights a clinical complication that may eventually develop as younger, lower-risk TAVR patients, age.

机构、总结、中文摘要均由AI分析而来,可能存在错误,请以论文正文为准。

↑