扩散加权SPECIAL改善超高磁场下J耦合代谢物的检测
Diffusion-weighted SPECIAL improves the detection of J-coupled metabolites at ultra-high magnetic field
AI总结:
提出DW-SPECIAL序列,通过结合半绝热SPECIAL与STE扩散模块缩短回波时间,显著减少J耦合代谢物信号损失,在14.1T下表现出优于STE-LASER的检测与扩散估计性能。
摘要:
提出了一种用于单体素扩散加权1H MRS(DWS)的新序列,命名为DW-SPECIAL,旨在改善强J耦合代谢物的检测及其后续扩散特性的估计。该序列将半绝热SPECIAL序列与受激回波(STE)扩散模块相结合。在14.1T下,对体模和大鼠脑体内进行了DW-SPECIAL和STE-LASER(目前高场啮齿动物DWS实验的金标准)的采集。对谷氨酸、谷氨酰胺(Gln)、肌醇、牛磺酸、总N-乙酰天冬氨酸、总胆碱、总肌酸和大分子物质的表观扩散系数和棒内扩散率(Callaghan模型)进行了拟合,并比较了两种序列的结果。DW-SPECIAL实现的更短回波时间(18 ms,而STE-LASER为33 ms)显著限制了由J演化引起的代谢物信号损失。此外,DW-SPECIAL保留了STE-LASER的主要优势:无交叉项、STE期间的扩散时间以及对B1不均匀性的有限敏感性。在体内,与STE-LASER相比,DW-SPECIAL产生了相同的光谱质量,并且无论b值如何,都降低了J耦合代谢物的Cramer-Rao下界(CRLB)。与基于平均衰减的拟合相比,DW-SPECIAL还降低了基于个体动物拟合的代谢物扩散估计的标准差,且没有损失准确性。我们得出结论,由于其缩短的回波时间,当研究像Gln这样的强J耦合代谢物时,DW-SPECIAL可作为STE-LASER的替代方案,从而在DWS采集背景下扩展了可及代谢物的范围。
英文摘要:
A new sequence for single-voxel diffusion-weighted 1H MRS (DWS), named DW-SPECIAL, is proposed to improve the detection and subsequent estimation of the diffusion properties of strongly J-coupled metabolites. It combines the semi-adiabatic SPECIAL sequence with a stimulated echo (STE) diffusion block. Acquisitions with DW-SPECIAL and STE-LASER, the current gold-standard for rodent DWS experiments at high fields, were performed at 14.1T on phantoms and in vivo on the rat brain. The apparent diffusion coefficient and intra-stick diffusivity (Callaghan's model) were fitted and compared between the sequences for glutamate, glutamine (Gln), myo-inositol, taurine, total N-acetylaspartate, total choline, total creatine and the macromolecules. The shorter echo time achieved with DW-SPECIAL (18 ms against 33 ms with STE-LASER) substantially limited the metabolites' signal loss caused by J-evolution. In addition, DW-SPECIAL preserved the main advantages of STE-LASER: absence of cross-terms, diffusion time during a STE and limited sensitivity to B1 inhomogeneities. In vivo, compared to STE-LASER, DW-SPECIAL yielded the same spectral quality and reduced the Cramer Rao Lower Bounds (CRLB) for J-coupled metabolites, irrespective of the b-value. DW-SPECIAL also reduced the standard deviation of the metabolites' diffusion estimates based on individual animal fitting without loss of accuracy compared to the fit on the averaged decay. We conclude that due to its reduced echo time, DW-SPECIAL can serve as an alternative to STE-LASER when strongly J-coupled metabolites like Gln are investigated, thereby extending the range of accessible metabolites in the context of DWS acquisitions.
机构、总结、中文摘要均由AI分析而来,可能存在错误,请以论文正文为准。