$^{177}$ Lu SPECT在存在$^{90}$ Y的情况下成像:$^{90}$ Y是否会降低图像定量?仿真研究
$^{177}$Lu SPECT Imaging in the Presence of $^{90}$Y: Does $^{90}$Y Degrade Image Quantification? A Simulation Study
摘要:
这项工作的目的是调查定量SPECT成像的准确性,$^{177}$ Lu的存在下,$^{90}$ Y,这发生在双同位素放射性药物治疗(RPT)涉及两种同位素。我们使用GATE蒙特卡罗模拟工具包进行体模研究,模拟填充有$^{177}$ Lu和$^{90}$ Y的球体放置在圆柱形水体模中,该圆柱形水体模也填充有两种放射性核素的活性。我们通过改变球体的位置、球体中$^{177}$ Lu和$^{90}$ Y的浓度以及背景活性的量来模拟多个体模配置和活性组合。我们研究了两种不同的散射窗口宽度与三重能量窗口(TEW)散射校正。我们还为每个配置创建了多个实现,以改进我们的评估,总共进行了540次模拟。使用模拟的Siemens SPECT相机对每种配置进行成像。使用标准3D OSEM算法重建投影,并确定与$^{177}$ Lu活性定量和对比噪声比(CNR)相关的误差。在所有配置中,定量误差在无$^{90}$ Y情况下的$\pm$ 6%内,并且我们发现,当存在$^{90}$ Y时,定量准确度可能略有提高,因为与TEW散射校正相关的误差减少。CNR不受$^{90}$ Y存在的显著影响,但当更宽的散射窗口宽度用于TEW散射校正时,CNR增加。散射窗的宽度对恢复的$^{177}$ Lu活性产生了1-2%的小但统计学显著的差异。基于这些结果,我们可以得出结论,$^{177}$ Lu的活性定量和病变可检测性不会因$^{90}$ Y的存在而降低。
英文摘要:
This work aims to investigate the accuracy of quantitative SPECT imaging of $^{177}$Lu in the presence of $^{90}$Y, which occurs in dual-isotope radiopharmaceutical therapy (RPT) involving both isotopes. We used the GATE Monte Carlo simulation toolkit to conduct a phantom study, simulating spheres filled with $^{177}$Lu and $^{90}$Y placed in a cylindrical water phantom that was also filled with activity of both radionuclides. We simulated multiple phantom configurations and activity combinations by varying the location of the spheres, the concentrations of $^{177}$Lu and $^{90}$Y in the spheres, and the amount of background activity. We investigated two different scatter window widths to be used with triple energy window (TEW) scatter correction. We also created multiple realizations of each configuration to improve our assessment, leading to a total of 540 simulations. Each configuration was imaged using a simulated Siemens SPECT camera. The projections were reconstructed using the standard 3D OSEM algorithm, and errors associated with $^{177}$Lu activity quantification and contrast-to-noise ratios (CNRs) were determined. In all configurations, the quantification error was within $\pm$6% of the no-$^{90}$Y case, and we found that quantitative accuracy may slightly improve when $^{90}$Y is present because of reduction of errors associated with TEW scatter correction. The CNRs were not significantly impacted by the presence of $^{90}$Y, but they were increased when a wider scatter window width was used for TEW scatter correction. The width of the scatter windows made a small but statistically significant difference of 1-2% on the recovered $^{177}$Lu activity. Based on these results, we can conclude that activity quantification of $^{177}$Lu and lesion detectability is not degraded by the presence of $^{90}$Y.
机构、总结、中文摘要均由AI分析而来,可能存在错误,请以论文正文为准。