非接触式声学微敲击光学相干弹性成像定量角膜各向异性弹性:在体兔研究
Non-contact acoustic micro-tapping optical coherence elastography for quantification of corneal anisotropic elasticity: in vivo rabbit study
摘要:
目的.证明使用非接触式和非侵入性光学相干弹性成像技术在体内准确测量角膜弹性模量。方法.采用非接触式动态声学微敲击光学相干弹性成像(AuT-OCE)技术,对兔角膜的弹性特性(面内杨氏模量E、面内剪切模量u和面外剪切模量G)进行了体内定量研究。在体内测量后,在浸提的整个球体中探索了测量的机械性能的IOP依赖性。一个几乎不可压缩的横观各向同性(NITI)模型被用来重建模量从AuT-OCE数据。独立地,角膜弹性模量也用传统的、破坏性的机械测试(拉伸伸长法和剪切流变法)离体测量。结果我们的研究表明,兔角膜弹性的各向异性很强。面内杨氏模量(计算为E=3u)在20-44 MPa的范围内,而面外剪切模量在34-261 kPa的范围内。在安乐死后1小时内对相同样本进行的压力依赖性离体OCE和破坏性机械试验均强烈支持AuT-OCE测量结果。结论.非接触式AuT-OCE可以在活体内无创地定量角膜各向异性弹性特性。翻译的相关性。由于OCT在眼科中被广泛接受,这些结果表明AuT-OCE快速转化为临床实践的潜力。此外,AuT-OCE可能会改善扩张性角膜疾病的诊断标准,从而实现早期诊断,减少并发症,定制手术治疗和个性化的眼睛生物力学模型。
英文摘要:
Purpose. To demonstrate accurate measurement of corneal elastic moduli in vivo with non-contact and non-invasive optical coherence elastography. Methods. Elastic properties (in-plane Young's modulus E and both in-plane, u, and out-of-plane, G, shear moduli) of rabbit cornea were quantified in vivo using non-contact dynamic Acoustic micro-Tapping Optical Coherence Elastography (AuT-OCE). The IOP-dependence of measured mechanical properties was explored in extracted whole globes following in vivo measurement. A nearly-incompressible transverse isotropic (NITI) model was used to reconstruct moduli from AuT-OCE data. Independently, cornea elastic moduli were also measured ex vivo with traditional, destructive mechanical tests (tensile extensometry and shear rheometry). Results. Our study demonstrates strong anisotropy of corneal elasticity in rabbits. The in-plane Young's modulus, computer as E=3u, was in the range of 20-44 MPa, whereas the out-of-plane shear modulus was in the range of 34-261 kPa. Both pressure-dependent ex vivo OCE and destructive mechanical tests performed on the same samples within an hour of euthanasia strongly support the results of AuT-OCE measurements. Conclusions. Non-contact AuT-OCE can non-invasively quantify cornea anisotropic elastic properties in vivo. Translational Relevance. As OCT is broadly accepted in Ophthalmology, these results suggest the potential for rapid translation of AuT-OCE into clinical practice. In addition, AuT-OCE can likely improve diagnostic criteria of ectatic corneal diseases, leading to early diagnosis, reduced complications, customized surgical treatment, and personalized biomechanical models of the eye.
机构、总结、中文摘要均由AI分析而来,可能存在错误,请以论文正文为准。