利用微型斩波器和标准中子监测器评估中子散射仪器的飞行时间、波长与距离校准方法

Evaluation of a method for time-of-flight, wavelength and distance calibration for neutron scattering instruments by means of a mini-chopper and standard neutron monitors

L. Vergara, M. Arai, R. Woracek, M. Olsson, A. Quintanilla, S. Alcock, J. Nilsson, K. Kanaki, P. M. Kadletz, O. Kirstein, R. Hall-Wilton, N. Tsapatsaris

AI总结:

研究评估利用微型斩波器和双位置标准中子监测器,由飞行时间校准中子波长与源距的方法,在V20光束线验证其精度并给出改进途径。

摘要:

将中子飞行时间(TOF)准确转换为波长及其不确定度,对中子散射测量具有根本重要性。尤其在仪器高度可配置的情况下,任何变更后都必须进行绝对波长的测定。受中子谱仪测定绝对波长方式的启发,据作者所知,我们首次评估了一种将TOF转换为中子波长的常用方法、监测器距中子源的距离,并解析计算了限制转换精度的不确定度贡献。该方法在Helmholtz Zentrum Berlin(HZB)的V20测试光束线进行评估,通过脉宽2.86 ms、重复频率14 Hz的脉冲模拟ESS源,并使用以140 Hz运行的微型斩波器、束流监测器(BMs)和数据采集基础设施。微型斩波器产生定义明确的中子脉冲,BM被放置在两个位置,从而能够测定所产生的每个脉冲的平均波长。所用实验装置在监测器位置实现了绝对波长测定,对于λ>4 Å,δλ_mean / λ_mean约为1.8%。通过适度增加参考监测器位置之间的距离,可实现低于0.5%的δλ_mean / λ_mean。使用更薄的监测器、更小的斩波器盘开口和更高转速的斩波器,还可进一步改进。该方法仅需两次中子测量,不需要使用晶体或复杂拟合,可作为散裂源上通常不使用快速斩波器的成像、衍射、反射仪和小角中子散射仪器的合适补充。

英文摘要:

Accurate conversion of neutron time-of-flight (TOF) to wavelength, and its uncertainty, is of fundamental importance to neutron scattering measurements. Especially in cases where instruments are highly configurable, the determination of the absolute wavelength after any change must always be performed. Inspired by the manner with which neutron spectrometers determine the absolute wavelength, we evaluate for the first time, in the author's knowledge, a commonly used method for converting TOF to neutron wavelength, the distance of a monitor from the source of neutrons and we analytically calculate the uncertainty contributions that limit the precision of the conversion. The method was evaluated at the V20 test beamline at the Helmholtz Zentrum Berlin (HZB), emulating the ESS source with a pulse of 2.86 ms length and 14 Hz repetition rate, by using a mini-chopper operated at 140 Hz, beam monitors (BMs) and data acquisition infrastructure. The mini-chopper created well-defined neutron pulses and the BM was placed at two positions, enabling the average wavelength of each of the pulses created to be determined. The used experimental setup resulted in absolute wavelength determination at the monitor positions with a $δλ_{mean} / λ_{mean}$ of $\sim$1.8% for $λ>4$ Å. With a modest increase of the distance between the reference monitor positions a $δλ_{mean} / λ_{mean}$ of below 0.5% can be achieved. Further improvements are possible by using a thinner monitor, smaller chopper disc openings and a higher rotational speed chopper. The method requires only two neutron measurements and doesn't necessitate the use of crystals or complex fitting, and could constitute a suitable addition to imaging, diffraction, reflectometers and small angle neutron scattering instruments, at spallation sources, that do not normally utilise fast choppers.

机构、总结、中文摘要均由AI分析而来,可能存在错误,请以论文正文为准。

↑