无精密设备的无线惯性测量组合的改进校准方法
Improved Calibration Procedure for Wireless Inertial Measurement Units without Precision Equipment
摘要:
惯性测量单元(IMU)在医疗应用中用于许多不同的目的。然而,IMU的测量精度会随着时间的推移而降低,从而需要重新校准。在2014年的论文中,Tedaldi等人提出了一种不需要外部精密设备或复杂程序的IMU校准方法。这允许没有惯性测量专业知识的最终用户或人员通过将传感器放置在几个合适但不精确定义的方向上来重新校准传感器。在这项工作中,我们提出了几个改进Tedaldi的方法,无论是在算法层面和校准过程:适应低噪声加速度计,校准助手对象,无线校准和丢包补偿。我们将修改后的校准程序应用于定制的IMU平台,并验证了多次校准运行结果的一致性。为了最大限度地减少重新校准所需的时间,我们分析了当使用较少的校准方向时校准结果准确度如何降低。我们发现,N=12个不同的方向足以实现非常好的校准,更多的方向只产生边际改善。与Tedaldi建议的37至50个方向相比,这是一个显著的改进。因此,我们减少了校准单个IMU所需的时间。5分钟至不到2分钟,而不会牺牲任何有意义的校准精度。
英文摘要:
Inertial measurement units (IMUs) are used in medical applications for many different purposes. However, an IMU's measurement accuracy can degrade over time, entailing re-calibration. In their 2014 paper, Tedaldi et al. presented an IMU calibration method that does not require external precision equipment or complex procedures. This allows end-users or personnel without expert knowledge of inertial measurement to re-calibrate the sensors by placing them in several suitable but not precisely defined orientations. In this work, we present several improvements to Tedaldi's method, both on the algorithmic level and the calibration procedure: adaptions for low noise accelerometers, a calibration helper object, and packet loss compensation for wireless calibration. We applied the modified calibration procedure to our custom-built IMU platform and verified the consistency of results across multiple calibration runs. In order to minimize the time needed for re-calibration, we analyzed how the calibration result accuracy degrades when fewer calibration orientations are used. We found that N=12 different orientations are sufficient to achieve a very good calibration, and more orientations yielded only marginal improvements. This is a significant improvement compared to the 37 to 50 orientations recommended by Tedaldi. Thus, we were reduced the time required to calibrate a single IMU from ca. 5 minutes to less than 2 minutes without sacrificing any meaningful calibration accuracy.
机构、总结、中文摘要均由AI分析而来,可能存在错误,请以论文正文为准。