用于PREX和CREX的精密Møller旋光仪
Precision Møller Polarimetry for PREX and CREX
摘要:
位于杰斐逊实验室A厅的PREX-2和CREX实验是对复杂原子核的宇称违反弹性电子散射的精确测量。一个要求是入射电子束偏振,通常为$\approx$ 90,以1%的精度已知。我们在光束线上调试并运行了一台Møller旋光仪,其精度超过了这一要求,PREX-2的精度达到了0.89%,CREX的精度达到了0.85%。不确定性是纯粹系统性的,从几个不同的来源积累,但由我们的知识的目标极化。我们的分析还表明,需要准确的原子波函数,以纠正Levchuk效应。我们描述的旋光仪操作和分析的细节,以及(CREX)从不同的旋光仪的基础上康普顿散射的结果进行比较。
英文摘要:
The PREX-2 and CREX experiments in Hall A at Jefferson Lab are precision measurements of parity violating elastic electron scattering from complex nuclei. One requirement was that the incident electron beam polarization, typically $\approx$90\%, be known with 1\% precision. We commissioned and operated a Møller polarimeter on the beam line that exceeds this requirement, achieving a precision of 0.89\% for PREX-2, and 0.85\% for CREX. The uncertainty is purely systematic, accumulated from several different sources, but dominated by our knowledge of the target polarization. Our analysis also demonstrates the need for accurate atomic wave functions in order to correct for the Levchuk Effect. We describe the details of the polarimeter operation and analysis, as well as (for CREX) a comparison to results from a different polarimeter based on Compton scattering.