COVID-19大流行期间使用的医疗和非医疗CPAP接口的性能评估
Performance assessment of medical and non-medical CPAP interfaces used during the COVID-19 pandemic
摘要:
背景资料:二零二零年初,意大利在短时间内出现大量COVID-19病例,导致医疗设备供应困难。为了解决这个问题,许多医疗保健操作人员重新调整了不同的口罩,但没有进行实验来评估它们的性能。本研究的目的是:评估三种面罩和一种CPAP头盔在其原始配置和修改后的性能,使用不同PEEP阀维持平均压力和半振幅变化,并分析抗菌(AB)或抗菌病毒(ABV)阀前PEEP过滤器对输送给患者的有效PEEP的影响。每个面罩上安装四个压力端口(头盔上三个),记录压力的平均值和半幅值。在PEEP阀之前使用任何AB、ABV过滤器进行测试。CPAP头盔是最有效的接口,产生更稳定的平均压力和不太突出的半振幅变化,但非医用面罩,特别是在修改后,保持稳定的平均压力值,只有适度增加的半振幅。与无过滤器条件相比,使用AB和ABV过滤器分别使平均压力增加2.23%和16.5%。CPAP头盔在检测性能方面是最可靠的接口,但重新适配的面罩可以确保几乎相似的性能。在PEEP阀之前使用ABV过滤器显著增加了检测到的平均压力,而AB过滤器几乎具有中性效果。
英文摘要:
Background: At the beginning of 2020, a high number of COVID-19 cases affected Italy in a short period, causing a difficult supply of medical equipment. To deal with the problem, many healthcare operators readapted different masks to medical devices, but no experiment was conducted to evaluate their performance. The aims of our study were: to assess the performances of three masks and a CPAP helmet in their original configuration and after modifications, in the maintenance of mean pressure and half-amplitude variations using different PEEP valves and to analyse the impact of antibacterial (AB) or antibacterial-viral (ABV) pre-valve PEEP filters on the effective PEEP delivered to the patients. Four pressure ports were installed on each mask (three on helmet), mean values and half amplitudes of pressure were recorded. Tests were performed with any, AB, ABV filter before the PEEP valve. CPAP helmet was the most efficient interface producing more stable mean pressure and less prominent half-amplitude variations but the non-medical masks, especially after the modifications, maintained a stable mean pressure value with only a moderate increase of half-amplitude. The use of AB and ABV filters, produced respectively an increase of 2,23% and 16.5% in mean pressure, compared to no filter condition. CPAP helmet is the most reliable interface in terms of detected performance, but readapted masks can assure almost a similar performance. The use of ABV filters before the PEEP valve significantly increases the detected mean pressure while the AB filters have almost a neutral effect.
机构、总结、中文摘要均由AI分析而来,可能存在错误,请以论文正文为准。