Photoplethysmographic Imaging (PPGI) allows the determination of pulse rate variability from sequential beat-to-beat intervals (BBI) and pulse wave velocity from spatially resolved recorded pulse waves. In either case, sufficient temporal accuracy is essential.
The presented work investigates the temporal accuracy of BBI estimation from photoplethysmographic signals. Within comprehensive numerical simulation, we systematically assess the impact of sampling rate, signal-to-noise ratio (SNR), and beat-to-beat shape variations on the root mean square error (RMSE) between real and estimated BBI.
Our results show that at sampling rates beyond 14 Hz only small errors exist when interpolation is used. For example, the average RMSE is 3 ms for a sampling rate of 14 Hz and an SNR of 18 dB. Further increasing the sampling rate only results in marginal improvements, e.g. more than tripling the sampling rate to 50 Hz reduces the error by approx. 14%. The most important finding relates to the SNR, which is shown to have a much stronger influence on the error than the sampling rate. For example, increasing the SNR from 18 dB to 24 dB at 14 Hz sampling rate reduced the error by almost 50% to 1.5 ms. Subtle beat-to-beat shape variations, moreover, increase the error decisively by up to 800%.
Our results are highly relevant in three regards: first, they partially explain different results in the literature on minimum sampling rates. Second, they emphasize the importance to consider SNR and possibly shape variation in investigations on the minimal sampling rate. Third, they underline the importance of appropriate processing techniques to increase SNR. Importantly, though our motivation is PPGI, the presented work immediately applies to contact PPG and PPG in other settings such as wearables. To enable further investigations, we make the scripts used in modelling and simulation freely available.
Beyond Single-Channel: Multichannel Signal Imaging for PPG-to-ECG Reconstruction with Vision Transformers
Department of Computer Science, University of Toronto(多伦多大学计算机科学系);Data Science Research Center, Duke Kunshan University(杜克昆山大学数据科学研究中心);Data Science Lab, Fields Institute for Research in Mathematical Sciences(数学科学研究院数据科学实验室);Department of Mathematics and Statistics, York University(约克大学数学与统计学系)
本文提出融合原始 PPG、一阶及二阶差分与曲线下面积的四通道信号图像表示,并以 ViT 建模心搏间和心搏内依赖,显著提升 PPG 到 ECG 的重建精度。