Signal-to-noise ratio is more important than sampling rate in beat-to-beat interval estimation from optical sensors

Sebastian Zaunseder, Antti Vehkaoja, Vincent Fleischhauer, Christoph Hoog Antink

摘要:

光电体积描记成像(PPGI)允许从连续的心跳到心跳间隔(BBI)和从空间分辨记录的脉搏波的脉搏波速度确定脉率变异性。在任何一种情况下,足够的时间精度是必不可少的。所提出的工作调查的时间精度BBI估计光电容积脉搏波信号。在全面的数值模拟中,我们系统地评估了采样率,信噪比(SNR),和节拍到节拍的形状变化的影响,对真实和估计BBI之间的均方根误差(RMSE)。我们的研究结果表明,在采样率超过14 Hz时,只有很小的误差存在插值。例如,对于14 Hz的采样率和18 dB的SNR,平均RMSE为3 ms。进一步增加采样率仅会导致边际改进,例如,将采样率增加三倍以上至50 Hz会将误差减少约10%。百分之十四最重要的发现涉及到的SNR,这被证明有一个更强的影响比采样率的错误。例如,在14 Hz采样率下将SNR从18 dB增加到24 dB,误差减少了近50%,降至1.5 ms。此外,细微的逐拍形状变化会使误差明显增加800%。我们的研究结果在三个方面是高度相关的:首先,他们部分解释了不同的结果在文献中的最小采样率。其次,他们强调了在最小采样率的调查中考虑SNR和可能的形状变化的重要性。第三,他们强调了适当的处理技术,以提高信噪比的重要性。重要的是,虽然我们的动机是PPGI,但所提出的工作立即适用于在其他设置(如可穿戴设备)中联系PPG和PPG。为了进一步研究,我们免费提供建模和模拟中使用的脚本。

英文摘要:

Photoplethysmographic Imaging (PPGI) allows the determination of pulse rate variability from sequential beat-to-beat intervals (BBI) and pulse wave velocity from spatially resolved recorded pulse waves. In either case, sufficient temporal accuracy is essential. The presented work investigates the temporal accuracy of BBI estimation from photoplethysmographic signals. Within comprehensive numerical simulation, we systematically assess the impact of sampling rate, signal-to-noise ratio (SNR), and beat-to-beat shape variations on the root mean square error (RMSE) between real and estimated BBI. Our results show that at sampling rates beyond 14 Hz only small errors exist when interpolation is used. For example, the average RMSE is 3 ms for a sampling rate of 14 Hz and an SNR of 18 dB. Further increasing the sampling rate only results in marginal improvements, e.g. more than tripling the sampling rate to 50 Hz reduces the error by approx. 14%. The most important finding relates to the SNR, which is shown to have a much stronger influence on the error than the sampling rate. For example, increasing the SNR from 18 dB to 24 dB at 14 Hz sampling rate reduced the error by almost 50% to 1.5 ms. Subtle beat-to-beat shape variations, moreover, increase the error decisively by up to 800%. Our results are highly relevant in three regards: first, they partially explain different results in the literature on minimum sampling rates. Second, they emphasize the importance to consider SNR and possibly shape variation in investigations on the minimal sampling rate. Third, they underline the importance of appropriate processing techniques to increase SNR. Importantly, though our motivation is PPGI, the presented work immediately applies to contact PPG and PPG in other settings such as wearables. To enable further investigations, we make the scripts used in modelling and simulation freely available.

机构、总结、中文摘要均由AI分析而来,可能存在错误,请以论文正文为准。

↑