Cherenkov成像系统提高放射治疗质量的回顾性评价

Retrospective Evaluation of an Always-on Cherenkov Imaging System for Radiotherapy Quality Improvement

Daniel A. Alexander, Michael Jermyn, Petr Bruza, Rongxiao Zhang, Erli Chen, Savannah M. Decker, Tatum L. McGlynn, Rory A. Rosselot, Jae Lee, Melanie L. Rose, Be… 展开作者

Daniel A. Alexander, Michael Jermyn, Petr Bruza, Rongxiao Zhang, Erli Chen, Savannah M. Decker, Tatum L. McGlynn, Rory A. Rosselot, Jae Lee, Melanie L. Rose, Benjamin B. Williams, Brian W. Pogue, David J. Gladstone, Lesley A. Jarvis

摘要:

目的:切伦科夫成像现在临床上可用于跟踪放射治疗的过程,作为治疗验证工具。这项工作的目的是通过对我们中心成像的回顾性审查,发现始终在线的切伦科夫图像作为一种新型事件检测和质量改进系统的好处。研究方法:在12个月期间,通过使用始终在线的商业系统自动采集切伦科夫成像,尝试对所有患者进行连续成像。多摄像机系统安装在我们中心放射肿瘤诊所的两个治疗掩体和附属卫星诊所的一个掩体中。作为正常治疗程序的一部分采集图像,并进行回顾性审查,标记潜在事件,由医生和医学物理团队进行评价。结果:作为本研究的一部分,共对622例患者进行了成像。在本总结中,确定了9例患者在治疗过程中发生的事件,仅通过切伦科夫成像检测到。在模拟、计划、治疗前审查和治疗输送过程中发现了与问题相关的事件,但在治疗输送之前未检测到任何事件。本研究中确定的主要改进领域是计划中非预期区域的剂量、由于定位导致的非预期区域的剂量以及设置期间的非理想团注放置。案例研究突出了这些问题的检测使用切伦科夫成像。结论:所有检测到的事件都被认为低于报告的阈值,但它们的观察可能会导致实践中的质量改进。也许最重要的是,成像是无缝的,不需要放疗团队的努力,并提供了实时和永久的记录,在每个部分提供了什么。

英文摘要:

Purpose: Cherenkov imaging is now clinically available to track the course of radiation therapy as a treatment verification tool. The aim of this work was to discover the benefits of always-on Cherenkov images as a novel incident detection and quality improvement system through retrospective review of imaging in our center. Methods: Continuous imaging of all patients was attempted during a 12-month period by automating the acquisition of Cherenkov imaging using an always-on commercial system. Multi-camera systems were installed in two treatment bunkers in the radiation oncology clinic at our center and one bunker in an affiliated satellite clinic. Images were acquired as part of normal treatment procedure and reviewed retrospectively with potential incidents flagged for evaluation by the physician and medical physics teams. Results: In total, 622 patients were imaged as part of this study. In this summary, 9 patients were identified with incidents occurring during their course of treatment that were detected only with Cherenkov imaging. Incidents were found relating to issues during simulation, planning, pre-treatment review, and treatment delivery, however none of the incidents were detected prior to treatment delivery. Primary areas of improvement identified in this study are dose to unintended areas in planning, dose to unintended areas due to positioning, and non-ideal bolus placement during setup. Case studies are presented highlighting the detection of these issues using Cherenkov imaging. Conclusions: All detected events were deemed below the threshold for reporting, but their observation could lead to quality improvement in practice. Perhaps most importantly, the imaging was seamless with no effort required by the radiotherapy team and provided both real-time and permanent records of what was delivered in each fraction.

机构、总结、中文摘要均由AI分析而来,可能存在错误,请以论文正文为准。

↑