用于硅光机原子力显微镜的甚高频探针
Very high frequency probes for atomic force microscopy with silicon optomechanics
摘要:
原子力显微镜(AFM)一直支持纳米科学和纳米技术超过30年,目前在许多领域,从凝聚态物理生物学。它能够在纳米尺度上测量非常微弱的力,从而阐明在基本过程中起作用的相互作用。在这里,我们利用微/纳机电系统和腔光学力学的综合优势,在100 MHz以上的频率制造动态模式AFM传感器。这比最快的商业AFM探针高出20年,这表明有机会在迄今为止尚未探索的时间尺度上测量力。制造是使用继承自光子硅电路的超大规模集成技术实现的。探头的环形光机械腔耦合到1.55 μ m激光,并具有130 MHz的机械谐振模式,空气中的品质因数为900。获得了3.10-16 m/sqrt(Hz)的位移检测极限,使得能够检测探针的布朗运动,并为具有皮米范围内的工作振幅的动态模式下的力感测实验铺平了道路。插入在一个定制的AFM仪器的实施例中,这种光机械传感器演示了执行力距离测量和保持针尖和样品之间的恒定的相互作用强度,AFM应用的基本要求的能力。实验表明,确实是一个稳定的闭环操作与4 nN/nm的一个前所未有的亚皮米振动振幅,其中尖端样品的相互作用是由拉伸的水弯月面介导的设定点。
英文摘要:
Atomic force microscopy (AFM) has been constantly supporting nanosciences and nanotechnologies for over 30 years, being present in many fields from condensed matter physics to biology. It enables measuring very weak forces at the nanoscale, thus elucidating interactions at play in fundamental processes. Here we leverage the combined benefits of micro/nanoelectromechanical systems and cavity optomechanics to fabricate a sensor for dynamic mode AFM at a frequency above 100 MHz. This is two decades above the fastest commercial AFM probes, suggesting opportunity for measuring forces at timescales unexplored so far. The fabrication is achieved using very-large scale integration technologies inherited from photonic silicon circuits. The probe's ring optomechanical cavity is coupled to a 1.55 um laser light and features a 130 MHz mechanical resonance mode with a quality factor of 900 in air. A limit of detection in displacement of 3.10-16 m/sqrt(Hz) is obtained, enabling the detection of the Brownian motion of the probe and paving the way for force sensing experiments in the dynamic mode with a working vibration amplitude in the picometer range. Inserted in a custom AFM instrument embodiment, this optomechanical sensor demonstrates the capacity to perform force-distance measurements and to maintain a constant interaction strength between tip and sample, an essential requirement for AFM applications. Experiments show indeed a stable closed-loop operation with a setpoint of 4 nN/nm for an unprecedented sub-picometer vibration amplitude, where the tip-sample interaction is mediated by a stretched water meniscus.
机构、总结、中文摘要均由AI分析而来,可能存在错误,请以论文正文为准。