arXiv 2108.10943eess.IVcs.CV

Correcting inter-scan motion artefacts in quantitative R1 mapping at 7T

Wellcome Centre for Human Neuroimaging, UCL Queen Square Institute of Neurology, University College London(伦敦大学学院皇后广场神经病学研究所威康人类神经影像中心);Athinoula A. Martinos Center for Biomedical Imaging, Massachusetts General Hospital and Harvard Medical School(麻省总医院与哈佛医学院阿蒂努拉·A·马丁诺斯生物医学影像中心);MR Research Collaborations, Siemens Healthcare Limited(西门子医疗有限公司磁共振研究合作部);Centre de Résonance Magnétique des Systèmes Biologiques, UMR5536, CNRS/University Bordeaux(法国国家科学研究中心/波尔多大学生物系统磁共振中心)

Yaël Balbastre, Ali Aghaeifar, Nadège Corbin, Mikael Brudfors, John Ashburner, Martina F. Callaghan

英文摘要:

Purpose: Inter-scan motion is a substantial source of error in $R_1$ estimation, and can be expected to increase at 7T where $B_1$ fields are more inhomogeneous. The established correction scheme does not translate to 7T since it requires a body coil reference. Here we introduce two alternatives that outperform the established method. Since they compute relative sensitivities they do not require body coil images. Theory: The proposed methods use coil-combined magnitude images to obtain the relative coil sensitivities. The first method efficiently computes the relative sensitivities via a simple ratio; the second by fitting a more sophisticated generative model. Methods: $R_1$ maps were computed using the variable flip angle (VFA) approach. Multiple datasets were acquired at 3T and 7T, with and without motion between the acquisition of the VFA volumes. $R_1$ maps were constructed without correction, with the proposed corrections, and (at 3T) with the previously established correction scheme. Results: At 3T, the proposed methods outperform the baseline method. Inter-scan motion artefacts were also reduced at 7T. However, reproducibility only converged on that of the no motion condition if position-specific transmit field effects were also incorporated. Conclusion: The proposed methods simplify inter-scan motion correction of $R_1$ maps and are applicable at both 3T and 7T, where a body coil is typically not available. The open-source code for all methods is made publicly available.

机构、总结、中文摘要均由AI分析而来,可能存在错误,请以论文正文为准。

↑