arXivDaily arXiv每日学术速递 周一至周五更新
arXiv 2103.07363quant-phphysics.ins-det
收藏

正弦波选通单光子探测器中SPAD噪声特性与门参数关系的研究

Investigation of the dependence of noise characteristics of SPAD on the gate parameters in sine-wave gated single-photon detectors

A. V. Losev, V. V. Zavodilenko, A. A. Koziy, Y. V. Kurochkin, A. A. Gorbatsevich

摘要:

本文研究了一种用于量子密钥分发(QKD)的1550 nm波长正弦波选通单光子探测器(SPD)。研究了不同门极参数对SPD噪声特性的影响。我们已经承认,随着门控电压的增加和量子效率(QE)的恒定值,暗计数率(DCR)下降。在此基础上提出了改进SPD和整个QKD器件性能的建议。发现当门控电压超过一定值时,DCR值迅速上升,且不同探测器的DCR值不同。它已被证明,通用的经验依赖编译连接SPD的控制和操作参数是一个不平凡的任务。

英文摘要:

In this paper, we have investigated a self-developed sine wave gated (SWG) single-photon detector (SPD) for 1550 nm wavelength primary for quantum key distribution (QKD) usage. We have investigated different gate parameters` influence on the SPD`s noise characteristics. We have admitted that with an increase of gating voltage and constant value of quantum efficiency (QE), the dark count rate (DCR) decreases. There have been made some recommendations to improve SPD`s and whole QKD device's characteristics based on these observations. There have been discovered the quick rise of the DCR value with the increase of gating voltage above some certain value and this value was different for different detectors. It has been shown that universal empirical dependence compilation to connect control and operational parameters of SPD is a non-trivial task.

机构、总结、中文摘要均由AI分析而来,可能存在错误,请以论文正文为准。

↑