This work evaluates the viability of polyimide and parylene-C for passivation of lithium-drifted silicon (Si(Li)) detectors. The passivated Si(Li) detectors will form the particle tracker and X-ray detector of the General Antiparticle Spectrometer (GAPS) experiment, a balloon-borne experiment optimized to detect cosmic antideuterons produced in dark matter annihilations or decays. Successful passivation coatings were achieved by thermally curing polyimides, and the optimized coatings form an excellent barrier against humidity and organic contamination. The passivated Si(Li) detectors deliver $\lesssim\,4$ keV energy resolution (FWHM) for 20$-$100 keV X-rays while operating at temperatures of $-$35 to $-45\,^{\circ}$C. This is the first reported successful passivation of Si(Li)-based X-ray detectors operated above cryogenic temperatures.
Models for light dark matter particles with masses below 1 GeV/c$^2$ are a natural and well-motivated alternative to so-far unobserved weakly interacting massive particles. Gram-scale cryogenic calorimeters provide the required detector performance to detect these particles and extend the direct dark matter search program of CRESST. A prototype 0.5 g sapphire detector developed for the $ν$-cleus experiment has achieved an energy threshold of $E_{th}=(19.7\pm 0.9)$ eV, which is one order of magnitude lower than previous results and independent of the type of particle interaction. The result pre…
There is a common need in astroparticle experiments such as direct dark matter detection, 0ν\b{eta}\b{eta} (double beta decay without emission of neutrinos) and Coherent Neutrino Nucleus Scattering experiments for light detectors with a very low energy threshold. By employing the Neganov-Luke Effect, the thermal signal of particle interactions in a semiconductor absorber operated at cryogenic temperatures, can be amplified by drifting the photogenerated electrons and holes in an electric field. This technology is not used in current experiments, in particular because of a reduction of the sign…
The bright scintillation of pure CsI operated at liquid-nitrogen temperature makes of this material a promising dark matter and neutrino detector. We present the first measurement of its quenching factor for nuclear recoils. Our findings indicate it is indistinguishable from that for sodium-doped CsI at room temperature. Additional properties such as light yield, afterglow, scintillation decay properties for electron and nuclear recoils, and energy proportionality are studied over the \mbox{108-165 K} temperature range, confirming the vast potential of this medium for rare-event searches.
We describe the design and in--flight performance of the cryostat and the self-contained $^{3}$He refrigerator for the OLIMPO balloon--borne experiment, a spectrophotometer to measure the Sunyaev-Zel'dovich effect in clusters of galaxies. The $^{3}$He refrigerator provides the 0.3 K operation temperature for the four arrays of kinetic inductance detectors working in 4 bands centered at 150, 250, 350 and 460 GHz. The cryostat provides the 1.65 K base temperature for the $^{3}$He refrigerator, and cools down the reimaging optics and the filters chain at about 2 K. The integrated system was desig…
Thermalization of a SQUID chip at cryogenic temperature: Thermal conductance measurement for GE 7031 Varnish Glue, Apiezon N Grease and Rubber Cement between 20 and 200 mK
针对 ATHENA X-IFU CryoAC 研发,本文在低于 20 mK 的稀释制冷机冷级上测量 2 mm × 2 mm SQUID 芯片经三种低温粘接材料固定后的热导率,并通过多次热循环考察退化效应。
RES-NOVA is a newly proposed experiment for the detection of neutrinos from astrophysical sources, mainly Supernovae, using an array of cryogenic detectors made of PbWO$_4$ crystals produced from archaeological Pb. This unconventional material, characterized by intrinsic high radiopurity, enables to achieve low-background levels in the region of interest for the neutrino detection via Coherent Elastic neutrino-Nucleus Scattering (CE$ν$NS). This signal lies at the detector energy threshold, O(1 keV), and it is expected to be hidden by naturally occurring radioactive contaminants of the crystal …
Mm and sub-mm waves polarimetry is the new frontier of research in Cosmic Microwave Background and Interstellar Dust studies. Polarimeters working in the IR to MM range need to be operated at cryogenic temperatures, to limit the systematic effects related to the emission of the polarization analyzer. In this paper we study the effect of the temperature of the different components of a waveplate polarimeter, and describe a system able to rotate, in a completely automated way, a birefringent crystal at 4K. We simulate the main systematic effects related to the temperature and non-ideality of the…
In this work, a first cryogenic characterization of a scintillating LiAlO$_{2}$ single crystal is presented. The results achieved show that this material holds great potential as a target for direct dark matter search experiments. Three different detector modules obtained from one crystal grown at the Leibniz-Institut für Kristallzüchtung (IKZ) have been tested to study different properties at cryogenic temperatures. Firstly, two 2.8 g twin crystals were used to build different detector modules which were operated in an above-ground laboratory at the Max Planck Institute for Physics (MPP) in M…
We present the design and characterization of a cryogenic phonon-sensitive 1-gram Si detector exploiting the Neganov-Trofimov-Luke effect to detect single-charge excitations. This device achieved 2.65(2)~eV phonon energy resolution when operated without a voltage bias across the crystal and a corresponding charge resolution of 0.03 electron-hole pairs at 100~V bias. With a continuous-readout data acquisition system and an offline optimum-filter trigger, we obtain a 9.2~eV threshold with a trigger rate of the order of 20~Hz. The detector's energy scale is calibrated up to 120~keV using an energ…
The cryogenic dark matter experiment CRESST-II aims at the direct detection of WIMPs via elastic scattering off nuclei in scintillating CaWO$_4$ crystals. We present a new, highly improved, detector design installed in the current run of CRESST-II Phase 2 with an efficient active rejection of surface-alpha backgrounds. Using CaWO$_4$ sticks instead of metal clamps to hold the target crystal, a detector housing with fully-scintillating inner surface could be realized. The presented detector (TUM40) provides an excellent threshold of ${\sim}\,0.60\,$keV and a resolution of $σ\,{\approx}\,0.090$ …
Current experiments aimed at measuring the polarization of the Cosmic Microwave Background (CMB) use cryogenic detector arrays and cold optical systems to boost the mapping speed of the sky survey. For these reasons, large volume cryogenic systems, with large optical windows, working continuously for years, are needed. Here we report on the cryogenic system of the QUBIC (Q and U Bolometric Interferometer for Cosmology) experiment: we describe its design, fabrication, experimental optimization and validation in the Technological Demonstrator configuration. The QUBIC cryogenic system is based on…
KAGRA is a second-generation interferometric gravitational-wave detector with 3-km arms constructed at Kamioka, Gifu in Japan. It is now in its final installation phase, which we call bKAGRA (baseline KAGRA), with scientific observations expected to begin in late 2019. One of the advantages of KAGRA is its underground location of at least 200 m below the ground surface, which brings small seismic motion at low frequencies and high stability of the detector. Another advantage is that it cools down the sapphire test mass mirrors to cryogenic temperatures to reduce thermal noise. In April-May 201…
The EDELWEISS collaboration has performed a search for dark matter particles with masses below the GeV-scale with a 33.4-g germanium cryogenic detector operated in a surface lab. The energy deposits were measured using a neutron-transmutation-doped Ge thermal sensor with a 17.7~eV (RMS) baseline heat energy resolution leading to a 60~eV analysis energy threshold. Despite a moderate lead shielding and the high-background environment, the first sub-GeV spin-independent dark matter limit based on a germanium target has been achieved. The experiment provides the most stringent, nuclear recoil base…
ATHENA is a large-class ESA mission selected for launch in 2031. It is designed to address the science theme "The Hot and Energetic Universe", performing X-ray observations (0.2-12 keV) at the L2 Sun-Earth Lagrangian point. The X-IFU is one of the two instruments of the payload. It is a cryogenic spectrometer providing spatially resolved high-resolution spectroscopy. The core of the instrument is a large array of TES microcalorimeters, operated at a 50 mK thermal bath. The X-IFU performances would be strongly degraded by the particle background expected in the L2 environment, thus advanced red…
The EDELWEISS collaboration has performed a search for Dark Matter (DM) particles interacting with electrons using a 33.4 g Ge cryogenic detector operated underground at the LSM. A charge resolution of 0.53 electron-hole pairs (RMS) has been achieved using the Neganov-Trofimov-Luke amplification with a bias of 78 V. We set the first Ge-based constraints on sub-MeV/c$^{2}$ DM particles interacting with electrons, as well as on dark photons down to 1 eV/c$^2$. These are competitive with other searches. In particular, new limits are set on the kinetic mixing of dark photon DM in a so far unconstr…
The CRESST experiment is a direct dark matter search which aims to measure interactions of potential dark matter particles in an earth-bound detector. With the current stage, CRESST-III, we focus on a low energy threshold for increased sensitivity towards light dark matter particles. In this manuscript we describe the analysis of one detector operated in the first run of CRESST-III (05/2016-02/2018) achieving a nuclear recoil threshold of 30.1eV. This result was obtained with a 23.6g CaWO$_4$ crystal operated as a cryogenic scintillating calorimeter in the CRESST setup at the Laboratori Nazion…
CRESST-II, standing for Cryogenic Rare Events Search with Superconducting Thermometers phase II, is an experiment searching for Dark Matter. In the LNGS facility in Gran Sasso, Italy, a cryogenic detector setup is operated in order to detect WIMPs by elastic scattering off nuclei, generating phononic lattice excitations and scintillation light. The thermometers used in the experiment consist of a tungsten thin-film structure evaporated onto the CaWO4 absorber crystal. The process of evaporation causes a decrease in the scintillation light output. This, together with the need of a big-scale det…