一种基于物理和学习的SPECT无传输衰减补偿方法
A physics and learning-based transmission-less attenuation compensation method for SPECT
摘要:
衰减补偿(AC)是可靠量化的先决条件,并且有利于单光子发射计算机断层成像(SPECT)中的视觉解释任务。典型的AC方法需要使用透射扫描(例如CT扫描)获得的衰减图。这有几个缺点,例如增加的辐射剂量,更高的成本,以及SPECT和CT扫描之间可能的不对准。此外,CT扫描通常不可用。在这种情况下,我们和其他人表明,散射光子在SPECT包含的信息,以估计衰减分布。为了利用这一观察结果,我们提出了一种基于物理和学习的方法,该方法使用SPECT发射数据在峰值和散射窗口中执行SPECT中的无透射AC。所提出的方法使用在散射窗口中获取的数据来使用基于物理的方法重建衰减图的初始估计。然后训练卷积神经网络将该初始估计分割成不同的区域。将预定义的衰减系数分配给这些区域,从而产生重建的衰减图,然后使用基于有序子集期望最大化的重建方法将其用于重建活动图。我们客观地评价了这种方法的性能,使用一个高度逼真的模拟研究进行临床相关的任务,检测心肌灌注SPECT灌注缺陷。我们的研究结果表明,使用所提出的方法和真正的衰减图实现的性能之间没有统计上的显着差异。从视觉上看,使用所提出的方法重建的图像看起来类似于真实的衰减图。总体而言,这些结果提供了所提出的方法执行无传输AC的能力的证据,并激励进一步的评估。
英文摘要:
Attenuation compensation (AC) is a pre-requisite for reliable quantification and beneficial for visual interpretation tasks in single-photon emission computed tomography (SPECT). Typical AC methods require the availability of an attenuation map obtained using a transmission scan, such as a CT scan. This has several disadvantages such as increased radiation dose, higher costs, and possible misalignment between SPECT and CT scans. Also, often a CT scan is unavailable. In this context, we and others are showing that scattered photons in SPECT contain information to estimate the attenuation distribution. To exploit this observation, we propose a physics and learning-based method that uses the SPECT emission data in the photopeak and scatter windows to perform transmission-less AC in SPECT. The proposed method uses data acquired in the scatter window to reconstruct an initial estimate of the attenuation map using a physics-based approach. A convolutional neural network is then trained to segment this initial estimate into different regions. Pre-defined attenuation coefficients are assigned to these regions, yielding the reconstructed attenuation map, which is then used to reconstruct the activity map using an ordered subsets expectation maximization-based reconstruction approach. We objectively evaluated the performance of this method using a highly realistic simulation study conducted on the clinically relevant task of detecting perfusion defects in myocardial perfusion SPECT. Our results showed no statistically significant differences between the performance achieved using the proposed method and that with the true attenuation maps. Visually, the images reconstructed using the proposed method looked similar to those with the true attenuation map. Overall, these results provide evidence of the capability of the proposed method to perform transmission-less AC and motivate further evaluation.
机构、总结、中文摘要均由AI分析而来,可能存在错误,请以论文正文为准。