束流损失监测中剂量测量用金刚石探测器的校准

Calibration of diamond detectors for dosimetry in beam-loss monitoring

G. Bassi, L. Bosisio, P. Cristaudo, M. Dorigo, A. Gabrielli, Y. Jin, C. La Licata, L. Lanceri, L. Vitale

摘要:

化学生长的金刚石晶体具有独特的性质,使其适合作为高辐射环境中的固态粒子探测器和剂量计。我们一直在使用基于化学气相沉积生长的单晶金刚石的传感器,用于SuperKEKB对撞机的剂量测定和光束损失监测。在这里,我们描述的组件和测试和校准程序套件,通过对该监测系统的金刚石探测器进行校准。我们报告了28个探测器上获得的结果,并评估这些设备的响应的稳定性和均匀性。

英文摘要:

Artificially-grown diamond crystals have unique properties that make them suitable as solid-state particle detectors and dosimeters in high-radiation environments. We have been using sensors based on single-crystal diamond grown by chemical vapour deposition for dosimetry and beam-loss monitoring at the SuperKEKB collider. Here we describe the assembly and the suite of test and calibration procedures adopted to characterise the diamond-based detectors of this monitoring system. We report the results obtained on 28 detectors and assess the stability and uniformity of response of these devices.