We present two different techniques for achieving low resistance ($<$20 n$\rm Ω$) contacts between copper and aluminium at cryogenic temperatures. The best method is based on gold plating of the surfaces in an e-beam evaporator immediately after Ar plasma etching in the same apparatus, yielding resistances as low as 3 n$\rm Ω$ that are stable over time. The second approach involves inserting indium in the Al/Cu joint. For both methods, we believe key elements are surface polishing, total removal of the aluminum oxide surface layer, and temporary application of large (typ. 11 kN) compression forces. We believe the values for gold plated contacts are the lowest ever reported for a Cu/Al joint of a few $\rm cm^{2}$. This technology could simplify the construction of thermal links for advanced cryogenics applications, in particular that of extremely low resistance heat switches for nuclear demagnetization refrigerators.
A new resistance bridge has been built at the Laboratoire national de métrologie et d'essais (LNE) to improve the ohm realization in the $Système International$ (SI) of units from the quantum Hall effect. We describe the instrument, the performance of which relies on two synchronized and noise-filtered current sources, an accurate and stable current divider and a cryogenic current comparator (CCC) having a low noise of $\mathrm{80~pA.t/Hz^{1/2}}$. The uncertainty budget for the measurement of the 100 $Ω/(R_\mathrm{K}/2)$ ratio, where $R_\mathrm{K}$ is the von Klitzing constant, amounts to a fe…
Helium-Gas-Cooled Cryogenic Current Comparator Integrated with a Quantum Resistance Standard
National Institute of Advanced Industrial Science and Technology (AIST), National Metrology Institute of Japan (NMIJ)(日本国立产业技术综合研究所(AIST),日本国家计量院(NMIJ))
Using a new brand of commercially available carbon resistor we built a cryogenic thermometer with an extremely good thermal contact to its thermal environment. Because of its superior thermal contact the thermometer is insensitive to low levels of spurious radio frequency heating. We calibrated our thermometer down to 5mK using a quartz tuning fork He-3 viscometer and measured its thermal resistance and thermal response time.
The X-IFU is the cryogenic spectrometer onboard the future ATHENA X-ray observatory. It is based on a large array of TES microcalorimeters, which works in combination with a Cryogenic AntiCoincidence detector (CryoAC). This is necessary to reduce the particle background level thus enabling part of the mission science goals. Here we present the first joint test of X-IFU TES array and CryoAC Demonstration Models, performed in a FDM setup. We show that it is possible to operate properly both detectors, and we provide a preliminary demonstration of the anti-coincidence capability of the system ach…
We implemented a low noise current preamplifier for the readout of resistive bolometers. We tested the apparatus on thermometer resistances ranging from 10 Ohm to 500 Mohm. The use of current preamplifier overcomes constraints introduced by the readout time constant due to the thermometer resistance and the input capacitance. Using cold JFETs, this preamplifier board is shown to have very low noise: the Johnson noise of the source resistor (1 fA/Hz1/2) dominated in our noise measurements. We also implemented a lock-in chain using this preamplifier. Because of fast risetime, compensation of the…
This paper presents a low noise amplifier for large arrays of silicon photomultipliers (SiPMs) operated in cryogenic environments, especially liquid argon (87 K) and liquid nitrogen (77 K). The goal is for one amplifier to read out a total photosensitive surface of tens of cm$^2$ while retaining the capability to resolve single photoelectron signals. Due to the large capacitance of SiPMs, typically a few nF per cm$^2$, the main contributor to noise is the series (voltage) component. A silicon-germanium heterojunction bipolar transistor (HBT) was selected as the input device of the cryogenic am…
We develop Johnson noise thermometry applicable to mesoscopic devices with variable source impedance with high bandwidth for fast data acquisition. By implementing differential noise measurement and two-stage impedance matching, we demonstrate noise measurement in the frequency range 120-250 MHz with a wide sample resistance range 30 Ω-100 kΩ tuned by gate voltages and temperature. We employ high-frequency, single-ended low noise amplifiers maintained at a constant cryogenic temperature in order to maintain the desired low noise temperature. We achieve thermometer calibration with temperature …
New inventions change the approach of vacuum suitable welding for research purpose. With orbital welding, laser welding and robot welding the possibilities increase to fabricate larger vessels more accurately. Despite this development there is still no perfect understanding on how to avoid virtual leaks and how to make such joints suitable for dynamic stress. By recalling its historical development, it is apparent how welding mistakes began occurring systematically and how to avoid them. With ASDEX-Upgrade as an example, it is shown how the attempt to conduct vacuum suitable welding has decrea…
A tabletop low-noise differential amplifer with a bandwidth of 100 kHz is presented. Low voltage drifts of the order of 100 nV/day are reached by thermally stabilizing relevant amplifer components. The input leakage current is below 100 fA. Input-stage errors are reduced by extensive circuitry. Voltage noise, current noise, input capacitance and input current are extraordinarily low. The input resistance is larger than 1 TOhm. The amplifers were tested with and deployed for electrical transport measurements of quantum devices at cryogenic temperatures.
We developed an impedance bridge that operates at cryogenic temperatures (down to 60 mK) and in perpendicular magnetic fields up to at least 12 T. This is achieved by mounting a GaAs HEMT amplifier perpendicular to a printed circuit board containing the device under test and thereby parallel to the magnetic field. The measured amplitude and phase of the output signal allows for the separation of the total impedance into an absolute capacitance and a resistance. Through a detailed noise characterization, we find that the best resolution is obtained when operating the HEMT amplifier at the highe…
The relative interfacial thermal contraction as a possible origin of the low-energy excess in cryogenic calorimeters
Max Planck Institute for Physics(马克斯·普朗克物理研究所);Marietta Blau Institute for Particle Physics of the Austrian Academy of Sciences(奥地利科学院玛丽埃塔·布劳粒子物理研究所);Humboldt-Universität zu Berlin(柏林洪堡大学)