两种不同的生物阻抗相移时间分辨测量装置的设计、研制和比较

Design, Development and Comparison of two Different Measurements Devices for Time-Resolved Determination of Phase Shifts of Bioimpedances

Roman Kusche, Steffen Kaufmann, Martin Ryschka

摘要:

生物阻抗测量是一种非侵入性的方法来确定有机组织的组成。为了测量两个电极之间的复生物阻抗,将具有恒定振幅的交流电注入组织中。所开发的电压降用于计算被测阻抗的实部和虚部。过去的测量结果表明,心脏跳动可能会对测试阻抗的实测相移产生影响。在这项工作中,开发了一个硬件系统,能够测量生物阻抗相移的变化与高分辨率。使用两种不同的测量方法。第一种方法是模拟电路,它已经在以前的项目中开发。另一种方法使用集成电路AFE 4300(Texas Instruments)[1]。两种方法的结果都用微控制器(来自Atmel Corp.的Xmega256A3BU)传输。通过USB接口连接到主机PC。在上位机上,利用开发的LabView程序实现了嵌入式系统的测量和控制的可视化。这项工作包括硬件和软件的开发,以及两种测量方法的精度比较。系统验证的结果是,两种测量方法的有效分辨率约为0.3°。

英文摘要:

Bioimpedance measurements are a non-invasive method to determine the composition of organic tissue. For measuring the complex bioimpedance between two electrodes, an alternating current with a constant amplitude is injected into the tissue. The developed voltage drop is used to calculate the real and imaginary part of the impedance under test. Measurements in the past indicated that it could be possible that the beating of the heart has an effect on the measured phase shift of the impedance under test. In this work a hardware system is developed, capable of measuring changes of the bioimpedance phase shift with high resolution. Two different measurement methods are used. The first method is an analog circuit, which has been developed in a previous project. The other method uses the integrated circuit AFE4300 (Texas Instruments) [1]. The results of both methods are transmitted with a microcontroller (Xmega256A3BU from Atmel Corp.) via the USB interface to a host PC. On the host PC the visualization of the measurements and the control of the embedded system is achieved with a developed LabView program. The work includes the development of the hardware and software, as well as a comparison of the accuracy of the two measurement methods. The result of a verification of the systems is that both measurement methods have an effective resolution of about 0.3°.

机构、总结、中文摘要均由AI分析而来,可能存在错误,请以论文正文为准。

↑