arXivDaily arXiv每日学术速递 周一至周五更新
arXiv 2003.07078physics.med-phphysics.app-ph

RF-coil with variable resonant frequency for multiheteronuclear ultra-high field MRI

V. A. Ivanov, A. A. Hurshkainen, G. A. Solomakha, M. A. Zubkov

英文摘要:

Here we propose double-coil setup to allow high signal-to-noise ratio broad-range heteronuclear magnetic resonance imaging experiments: two independent coils, one of them tuned to $^{1}$H frequency to perform anatomical $^{1}$H imaging, and another one, metamaterial-inspired coil, tuned to the X-nucleus frequency. In this work our goal was to design a broad-range X-nuclei coil to cover $^{2}$H, $^{11}$B, $^{13}$C, $^{23}$Na, $^{7}$Li and $^{31}$P nuclear magnetic resonance frequencies, and to combine it with $^{1}$H coil in one setup. The system was designed for 11.7 T scanner, i.e., with 76-203 MHz frequency tuning range for the X-nuclei and tuned to 500 MHz for the proton coil. X-nuclei coil operates via excitation of the fundamental eigenmode of an array of parallel non-magnetic wires. The excitation of the array is provided via non-resonant feeding loop inductively coupled to the resonator. In order to tune the X-coil over such a wide range, both structural capacitance and inductance of the coil were made variable; narrow range tuning of the $^{1}$H coil is achieved via conventional tuning-matching circuit. Here, the design principle and setup tunability were investigated in simulations and experimentally.

机构、总结、中文摘要均由AI分析而来,可能存在错误,请以论文正文为准。

↑