利用一次活塞式压力计实现一次测温ppm级压力稳定
Realization of ppm level pressure stability for primary thermometry using a primary piston gauge
摘要:
为了使单压折射率气体测温法(SPRIGT)的不确定度达到0.25 mK,He-4气体在30 kPa至90 kPa范围内的相对压力变化不应超过4 ppm(k=1)。为此,开发了一种新型的压力控制系统。它由两个主要部分组成:控制压力的活塞式压力计和用于补充活塞式压力计微泄漏的自制气体补偿系统。此外,为了保持活塞在恒定的高度,使用伺服回路,该伺服回路实时自动确定所需的额外气体量。在室温下,稳定压力的标准偏差在30 kPa时为3.0 mPa,在60 kPa时为4.5 mPa,在90 kPa时为2 mPa。在5 K-25 K的温度范围内,SPRIGT的相对压力稳定性优于0.16 ppm,即比要求的稳定性好25倍。此外,相同的压力稳定系统可容易地转换到其他主气体温度计。
英文摘要:
To achieve an uncertainty of 0.25 mK in single-pressure refractive-index gas thermometry (SPRIGT), the relative pressure variation of He-4 gas in the range 30 kPa to 90 kPa, should not exceed 4 ppm (k=1). To this end, a novel pressure control system has been developed. It consists of two main parts: a piston gauge to control the pressure, and a home-made gas compensation system to supplement the micro-leak of the piston gauge. In addition, to maintain the piston at constant height, a servo loop is used that automatically determines in real time the amount of extra gas required. At room temperature, the standard deviations of the stabilized pressure are 3.0 mPa at 30 kPa, 4.5 mPa at 60 kPa and 2 mPa at 90 kPa. For the temperature region 5 K-25 K used for SPRIGT in the present work, the relative pressure stability is better than 0.16 ppm i.e. 25 times better than required. Moreover, the same pressure stabilization system is readily transposable to other primary gas thermometers.
机构、总结、中文摘要均由AI分析而来,可能存在错误,请以论文正文为准。