气体平衡气体调制折射法(GEq-GAMOR),用于压力评估,精度低于ppm
Gas equilibration gas modulation refractometry (GEq-GAMOR) for assessment of pressure with sub-ppm precision
摘要:
提出了一种新的气体调制折射率测量方法(GAMOR),其性能优于原方法[单腔调制GAMOR(SCM-GAMOR)]。通过平衡两个腔中的压力来进行参考测量。这样,达到参考测量的适当条件所花费的时间已经减少。这意味着测量周期的更大部分可以用于数据采集,从而减少白噪声并改善短期特性。所提出的实现还包括具有改进的温度稳定性和评估的新腔设计。这有助于改善长期特性。该系统的特点是相对于一个自重活塞规。对于较短的积分时间(最多10分钟),它可以提供超过原始SCM-GAMOR系统两倍的响应。对于更长的积分时间,并且长达18小时,系统在4303 Pa下显示出1 mPa的积分时间无关的Allan偏差(对应于定义为Allan偏差的两倍的精度,0.5 ppm)。这意味着,对于所有的积分时间,新系统相对于原始实现显示出显著的改进(对于18小时的积分时间,系数为8)。当用于低压时,它可以提供亚mPa区域的精度;对于具有抽空测量腔的情况,系统提供的精度高达约1000 mPa。40个测量周期(ca. 1.5小时),0.7 mPa sqrt(周期)的白噪声限制噪声,以及0.15 mPa的最小Allan偏差。此外,在所研究的压力范围内,即在2.8 - 10.1 kPa范围内,它显示出相对于自重活塞压力计的纯线性响应。这意味着该系统可用于在大压力范围内传递校准,具有极低的不确定性。
英文摘要:
A novel realization of Gas Modulation Refractometry (GAMOR) that outperforms the original realization [Single Cavity Modulated GAMOR (SCM-GAMOR)], is presented. The reference measurements are carried out by equalizing the pressures in the two cavities. By this, the time it takes to reach adequate conditions for the reference measurements has been reduced. This implies that a larger fraction of the measurement cycle can be devoted to data acquisition, which reduces white noise and improves on short-term characteristics. The presented realization also encompasses a new cavity design with improved temperature stabilization and assessment. This has contributed to an improved long-term characteristics. The system was characterized with respect to a dead weight piston gauge. For short integration times (up to 10 min) it can provide a response that exceeds that of the original SCM-GAMOR system by a factor of two. For longer integration times, and up to 18 hours, the system shows, at 4303 Pa, an integration time independent Allan deviation of 1 mPa (corresponding to a precision defined as twice the Allan deviation, of 0.5 ppm). This implies that the novel system shows a significant improvement with respect to the original realization for all integration times (by a factor of 8 for an integration time of 18 hours). When used for low pressures, it can provide a precision in the sub-mPa region; for the case with an evacuated measurement cavity, the system provided, up to ca. 40 measurement cycles (ca. 1.5 hours), a white-noise limited noise of 0.7 mPa sqrt(cycle), and minimum Allan deviation of 0.15 mPa. Furthermore, over the pressure range investigated, i.e. in the 2.8 - 10.1 kPa range, it shows, with respect to a dead weight piston gauge, a purely linear response. This implies that the system can be used for transfer of calibration over large pressure ranges with exceptional low uncertainty.
机构、总结、中文摘要均由AI分析而来,可能存在错误,请以论文正文为准。