This report presents an overview of the unconventional use of charge-coupled devices (CCDs) to search for Dark Matter (DM). The DArk Matter In CCDs (DAMIC Experiment) employs the bulk silicon of thick, fully-depleted CCDs as a target for ionization signals produced by interactions of particle dark matter from the galactic halo. The DAMIC collaboration has engaged in an extensive campaign of characterization efforts to understand the response of these CCDs to low-energy nuclear recoils and their unique capabilities, including the use of high spatial resolution for both the rejection and study of backgrounds. The preliminary results of DAMIC prove the performance of the detector, provide measurements of the background contamination and demonstrate the potentiality for DM searches, with only ~40 grams of detector mass. The next phase of the experiment, DAMIC-M (DArk Matter in CCDs at Modane), will consist of a kg-sized detector, implementing the most massive CCDs ever built. These CCDs will feature sub-electron noise and will be deployed in a low-radioactivity environment at the Laboratoire Souterrain de Modane in France.
Dark Matter In CCDs (DAMIC) is a silicon detector apparatus used primarily for searching for low-mass dark matter using the silicon bulk of Charge-Coupled Devices (CCDs) as targets. The silicon target within each CCD is \SI{675}{\micro\meter} thick and its top surface is divided into over 16 million \SI{15}{\micro\meter} $\times$ \SI{15}{\micro\meter} pixels. The DAMIC collaboration has installed a number of these CCDs at SNOLAB. As of 2019, DAMIC at SNOLAB has reached operational conditions with leakage current less than \SI{8.2e-22}{\ampere\per\centi\meter\squared} and a readout noise of \SI…
The DAMIC (Dark Matter in CCDs) experiment searches for the interactions of dark matter particles with the nuclei and the electrons in the silicon bulk of thick fully depleted charge-coupled devices (CCDs). Because of the low noise and low dark current, DAMIC CCDs are sensitive to the ionization signals expected from low-mass dark matter particles ($< 10$ GeV). A 40-gram target detector has collected data at the SNOLAB underground laboratory since 2017. Recent results from the searches for DM-electron scattering and hidden-photon absorption will be summarized and the status of WIMPs-nucleon se…
We present measurements of bulk radiocontaminants in the high-resistivity silicon CCDs from the DAMIC at SNOLAB experiment. We utilize the exquisite spatial resolution of CCDs to discriminate between $α$ and $β$ decays, and to search with high efficiency for the spatially-correlated decays of various radioisotope sequences. Using spatially-correlated $β$ decays, we measure a bulk radioactive contamination of $^{32}$Si in the CCDs of $140 \pm 30$ $μ$Bq/kg, and place an upper limit on bulk $^{210}$Pb of $< 160~μ$Bq/kg. Using similar analyses of spatially-correlated bulk $α$ decays, we set limits…
We present measurements of radioactive contamination in the high-resistivity silicon charge-coupled devices (CCDs) used by the DAMIC experiment to search for dark matter particles. Novel analysis methods, which exploit the unique spatial resolution of CCDs, were developed to identify $α$ and $β$ particles. Uranium and thorium contamination in the CCD bulk was measured through $α$ spectroscopy, with an upper limit on the $^{238}$U ($^{232}$Th) decay rate of 5 (15) kg$^{-1}$ d$^{-1}$ at 95% CL. We also searched for pairs of spatially correlated electron tracks separated in time by up to tens of …
The DAMIC experiment employs large-area, thick charge-coupled devices (CCDs) to search for the interactions of low-mass dark matter particles in the galactic halo with silicon atoms in the CCD target. From 2017 to 2019, DAMIC collected data with a seven-CCD array (40-gram target) installed in the SNOLAB underground laboratory. We report dark-matter search results, including a conspicuous excess of events above the background model below 200 eV$_{\rm ee}$, whose origin remains unknown. We present details of the published spectral analysis, and update on the deployment of skipper CCDs to perform…
The DAMIC experiment uses fully depleted, high resistivity CCDs to search for dark matter particles. With an energy threshold $\sim$50 eV$_{ee}$, and excellent energy and spatial resolutions, the DAMIC CCDs are well-suited to identify and suppress radioactive backgrounds, having an unrivaled sensitivity to WIMPs with masses $<$6 GeV/$c^2$. Early results motivated the construction of a 100 g detector, DAMIC100, currently being installed at SNOLAB. This contribution discusses the installation progress, new calibration efforts near the threshold, a preliminary result with 2014 data, and the prosp…
The DAMIC (Dark Matter in CCDs) experiment uses high resistivity, scientific grade CCDs to search for dark matter. The CCD's low electronic noise allows an unprecedently low energy threshold of a few tens of eV that make it possible to detect silicon recoils resulting from interactions of low mass WIMPs. In addition the CCD's high spatial resolution and the excellent energy response results in very effective background identification techniques. The experiment has a unique sensitivity to dark matter particles with masses below 10 GeV/c$^2$. Previous results have demonstrated the potential of t…
The DAMIC-M (DArk Matter In CCDs at Modane) experiment employs thick, fully depleted silicon charged-coupled devices (CCDs) to search for dark matter particles with a target exposure of 1 kg-year. A novel skipper readout implemented in the CCDs provides single electron resolution through multiple non-destructive measurements of the individual pixel charge, pushing the detection threshold to the eV-scale. DAMIC-M will advance by several orders of magnitude the exploration of the dark matter particle hypothesis, in particular of candidates pertaining to the so-called "hidden sector." A prototype…
Experimental efforts of the last decades have been unsuccessful in detecting WIMPs (Weakly Interacting Massive Particles) in the 10-to-10$^4$ GeV/$c^2$ range, thus motivating the search for lighter dark matter. The DAMIC (DArk Matter In CCDs) at SNOLAB experiment aims for direct detection of light dark matter particles ($m_χ<$ 10 GeV/$c^2$) by means of CCDs (Charge-Coupled Devices). Fully-depleted 675 $μ$m-thick CCDs are used to such end. The optimized readout noise and operation at cryogenic temperatures allow for a detection threshold of 50 eV$_{\text{ee}}$ electron-equivalent energy. Focusi…
We construct the first comprehensive radioactive background model for a dark matter search with charge-coupled devices (CCDs). We leverage the well-characterized depth and energy resolution of the DAMIC at SNOLAB detector and a detailed GEANT4-based particle-transport simulation to model both bulk and surface backgrounds from natural radioactivity down to 50 eV$_{\text{ee}}$. We fit to the energy and depth distributions of the observed ionization events to differentiate and constrain possible background sources, for example, bulk $^{3}$H from silicon cosmogenic activation and surface $^{210}$P…
Nuclear recoils in crystal detectors generate radiation damage in the form of crystal defects that can be measured in scientific-grade CCDs as local hot spots of leakage current stimulated by temperature increases in the devices. In this proceeding, we use a neutron source to generate defects in DAMIC-M CCDs, and using increases in leakage current at different temperatures, we demonstrate a procedure for identifying crystal defects in the CCDs of the DAMIC-M experiment. This is the first time that individual defects generated from nuclear recoils have been studied. This technique could be used…
We introduce the fully-depleted charge-coupled device (CCD) as a particle detector. We demonstrate its low energy threshold operation, capable of detecting ionizing energy depositions in a single pixel down to 50 eVee. We present results of energy calibrations from 0.3 keVee to 60 keVee, showing that the CCD is a fully active detector with uniform energy response throughout the silicon target, good resolution (Fano ~0.16), and remarkable linear response to electron energy depositions. We show the capability of the CCD to localize the depth of particle interactions within the silicon target. We…
DAMIC (Dark Matter in CCDs) is a novel dark matter experiment that has unique sensitivity to dark matter particles with masses below 10 GeV. Due to its low electronic readout noise (R.M.S. ~3 e-) this instrument is able to reach a detection threshold below 0.5 keV nuclear recoil energy, making the search for dark matter particles with low masses possible. We report on early results and experience gained from a detector that has been running at SNOLAB from Dec 2012. We also discuss the measured and expected backgrounds and present the plan for future detectors to be installed in 2014.
In this work, we will present a physical model and measurements of the transport of small charge packets in the bulk of thick high resistivity CCD before being collected by the pixel potential wells. A new technique to measure the lateral spread of the charge as a function of the ionization depth in the bulk is presented. Results from measurements on CCD currently in use for several scientific instruments are shown and validated with a new mathematical algorithm to extend the current modeling based only on the diffusion of the charge in silicon.
We characterize the astrometric distortion at the edges of thick, fully-depleted CCDs in the lab using a bench-top simulation of LSST observing. By illuminating an array of forty thousand pinholes (30mu m diameter) at the object plane of a f/1.2 optical reimager, thousands of PSFs can be imaged over a 4Kx4K pixel CCD. Each high purity silicon pixel, 10mu m square by 100mu m deep, can then be individually characterized through a series of sub-pixel dithers in the X/Y plane. The unique character [response, position, shape] of each pixel as a function of flux, wavelength, back side bias, etc. can…
Searches for CEνNS and Physics beyond the Standard Model using Skipper-CCDs at CONNIE
Instituto de Ciencias Nucleares, Universidad Nacional Autónoma de México(墨西哥国立自治大学核科学研究所);Centro Atómico Bariloche and Instituto Balseiro, Comisión Nacional de Energía Atómica (CNEA), Consejo Nacional de Investigaciones Científicas y Técnicas (CONICET), Universidad Nacional de Cuyo (UNCUYO)(巴里洛切原子中心与巴尔塞罗研究所,阿根廷国家原子能委员会(CNEA),阿根廷国家科学研究委员会(CONICET),门多萨国立大学(UNCUYO));International Center for Advanced Studies & Instituto de Ciencias Físicas, ECyT-UNSAM and CONICET(高级研究中心与物理科学研究所,阿根廷国立萨米恩托理工大学及阿根廷国家科学研究委员会(CONICET));Instituto de Física, Universidade Federal do Rio de Janeiro(里约热内卢联邦大学物理研究所);Fermi National Accelerator Laboratory(费米国家加速器实验室)
Low noise CCDs fully-depleted up to 675 micrometers have been identified as a unique tool for Dark Matter searches and low energy neutrino physics. The charge collection efficiency (CCE) for these detectors is a critical parameter for the performance of future experiments. We present here a new technique to characterize CCE in back-illuminated CCDs based on soft X-rays. This technique is used to characterize two different detector designs. The results demonstrate the importance of the backside processing for detection near threshold, showing that a recombination layer of a few microns signific…
In this paper we describe the Dark Energy Camera (DECam), which will be the primary instrument used in the Dark Energy Survey. DECam will be a 3 sq. deg. mosaic camera mounted at the prime focus of the Blanco 4m telescope at the Cerro-Tololo International Observatory (CTIO). It consists of a large mosaic CCD focal plane, a five element optical corrector, five filters (g,r,i,z,Y), a modern data acquisition and control system and the associated infrastructure for operation in the prime focus cage. The focal plane includes of 62 2K x 4K CCD modules (0.27"/pixel) arranged in a hexagon inscribed wi…
We report a measurement of the ionization efficiency of silicon nuclei recoiling with sub-keV kinetic energy in the bulk silicon of a charge-coupled device (CCD). Nuclear recoils are produced by low-energy neutrons ($<$24 keV) from a $^{124}$Sb-$^{9}$Be photoneutron source, and their ionization signal is measured down to 60 eV electron equivalent. This energy range, previously unexplored, is relevant for the detection of low-mass dark matter particles. The measured efficiency is found to deviate from the extrapolation to low energies of the Lindhard model. This measurement also demonstrates th…
Fully-depleted thick silicon Skipper-charge-coupled devices (Skipper-CCDs) are an important technology to probe neutrino and light-dark-matter interactions due to their sub-electron read-out noise. However, the successful search for rare neutrino or dark-matter events requires the signal and all backgrounds to be fully characterized. In particular, a measurement of the electron-hole pair creation energy below 150 eV and the Fano factor are necessary for characterizing the dark matter and neutrino signals. Moreover, photons from background radiation may Compton scatter in the silicon bulk, prod…
We present the first direct-detection search for eV-to-GeV dark matter using a new ~2-gram high-resistivity Skipper-CCD from a dedicated fabrication batch that was optimized for dark-matter searches. Using 24 days of data acquired in the MINOS cavern at the Fermi National Accelerator Laboratory, we measure the lowest rates in silicon detectors of events containing one, two, three, or four electrons, and achieve world-leading sensitivity for a large range of sub-GeV dark matter masses. Data taken with different thicknesses of the detector shield suggest a correlation between the rate of high-en…