压阻骨水泥与电阻抗断层成像在全关节置换中的界面载荷监测和失效检测

Interfacial Load Monitoring and Failure Detection in Total Joint Replacements via Piezoresistive Bone Cement and Electrical Impedance Tomography

Hamid Ghaednia, Crystal E. Owens, Ricardo Roberts, Tyler N. Tallman, A. John Hart, Kartik M. Varadarajan

摘要:

无菌性松动或植入物固定丢失是全关节置换术后的常见并发症。在美国,翻修手术每年花费医疗保健系统超过80亿美元。尽管无菌性松动普遍存在,但及时准确地检测仍然是一个挑战,因为传统的成像方式(如平片)难以可靠地检测早期植入物松动。基于这一挑战,我们提出了一种新的方法,用于骨水泥关节置换的体内监测和失效检测。聚(甲基丙烯酸甲酯)(PMMA)骨水泥改性与低体积分数的短切碳纤维(CF),赋予压电为基础的自传感。电阻抗断层扫描(EIT),然后用于检测和监测负载引起的变形和断裂的CF/PMMA在一个幻影坦克。因此,我们表明,EIT确实能够熟练地检测假体替代物上的负载力,区分增加的负载大小,检测植入物固定失败,甚至区分骨水泥开裂和骨水泥脱粘。由于EIT是一种低成本、生理良性且可能实时成像的方式,因此本文所述的可行性研究有可能改变植入物状况术后监测的护理标准。除了临床相关性之外,该技术还可以通过体内监测深入了解引发无菌性松动的因素,对骨科研究人员产生积极影响。

英文摘要:

Aseptic loosening, or loss of implant fixation, is a common complication following total joint replacement. Revision surgeries cost the healthcare system over $8 billion annually in the US. Despite the prevalence of aseptic loosening, timely and accurate detection remains a challenge because traditional imaging modalities such as plain radiographs struggle to reliably detect the early stages of implant loosening. Motivated by this challenge, we present a novel approach for in vivo monitoring and failure detection of cemented joint replacements. Poly(methyl methacrylate) (PMMA) bone cement is modified with low volume fractions of chopped carbon fiber (CF) to impart piezoresistive-based self-sensing. Electrical impedance tomography (EIT) is then used to detect and monitor load-induced deformation and fracture of CF/PMMA in a phantom tank. We therefore show that EIT indeed is able to adeptly detect loading force on a prosthetic surrogate, distinguish between increasing load magnitudes, detect failure of implant fixation, and even distinguish between cement cracking and cement de-bonding. Because EIT is a low-cost, physiologically benign, and potentially real-time imaging modality, the feasibility study herein presented has potential to transform the standard of care for post-operative monitoring of implant conditions. Beyond clinical relevance, this technique could positively impact orthopedic researchers by providing, via in vivo monitoring, insight into the factors that initiate aseptic loosening.

机构、总结、中文摘要均由AI分析而来,可能存在错误,请以论文正文为准。

↑