低导热材料热盘瞬态平面源法的分析与改进
Analysis and Improvement of the Hot Disk Transient Plane Source Method for Low Thermal Conductivity Materials
摘要:
热盘瞬态平面源(TPS)方法是一种广泛使用的标准技术(ISO 22007-2),用于表征材料的热性能,特别是热导率,k。尽管热盘TPS方法对于各种普通材料具有良好的可靠性,但是当应用于低k热绝缘材料时,还已知热盘TPS方法遭受显著的系统误差。在这里,我们提出了一个相结合的数值和实验研究的几何形状的热盘传感器的测量值的低k材料的影响。我们表明,误差是强烈影响的有限厚度和热质量的传感器的绝缘层,以及相应的增加的有效加热器的大小超出了半径的嵌入式金属加热器本身。我们还数值研究了误差对样品热性能的依赖性,证实了低k样品的误差更严重。还提供了一个简单的校正功能,将标准热盘TPS测量的明显(错误)结果转换为更准确的值。还使用湿法和干法蚀刻对标准聚酰亚胺传感器进行了优化,以直接提供更准确的测量。使用基于标准TPS传感器导出的数值校正因子对Airloy x56气凝胶和商业二氧化硅气凝胶的k的实验校正值与使用优化的聚酰亚胺传感器的TPS传感器的直接测量值非常一致。这两种方法都可以将误差降低到小于4%,而原始传感器测量的原始值的高估误差约为40%。这样的结果表明,对原始传感器或优化传感器的数值校正都能够为这样的材料提供高度准确的热导率值。
英文摘要:
The hot disk transient plane source (TPS) method is a widely used standard technique (ISO 22007-2) for the characterization of thermal properties of materials, especially the thermal conductivity, k. Despite its well-established reliability for a wide variety of common materials, the hot disk TPS method is also known to suffer from a substantial systematic errors when applied to low-k thermal insulation materials. Here, we present a combined numerical and experimental study on the influence of the geometry of hot disk sensor on measured value of low-k materials. We demonstrate that the error is strongly affected by the finite thickness and thermal mass of the sensor's insulation layer was well as the corresponding increase of the effective heater size beyond the radius of the embedded metal heater itself. We also numerically investigate the dependence of the error on the sample thermal properties, confirming that the errors are worse in low-k samples. A simple correction function is also provided, which converts the apparent (erroneous) result from a standard hot disk TPS measurement to a more accurate value. A standard polyimide sensor was also optimized using both wet and dry etching to provide more accurate measurement directly. Experimentally corrected value of k for Airloy x56 aerogel and a commercial silica aerogel using the numerical correction factor derived based on the standard TPS sensor is in excellent agreement with the directly measured value from the TPS sensor using the optimized polyimide sensor. Both of these methods can reduce the errors to less than 4% as compared to around 40% error of overestimation from raw values measured with the pristine sensor. Such results show that both the numerical correction to a pristine senor or an optimized sensor are capable of providing highly accurate value of thermal conductivity for such materials.
机构、总结、中文摘要均由AI分析而来,可能存在错误,请以论文正文为准。