R1标测方案的比较:我们测量的是什么?
Comparison of R1 Mapping Protocols: What are we measuring?
摘要:
目的:最近的工作强调了个别组织的自旋晶格弛豫率($R_1$)的广度。一个潜在的变异来源是用于测定$R_1$的方案。研究了R1和弛豫率$r_1$的方法依赖性。研究方法:在具有不同浓度的MnCl 2的凝胶模型中定量$R_1$,并且使用不同的采集方法对三名健康志愿者的小队列进行定量。将Siemens反转恢复(IR)和饱和恢复(SR)方案应用于体模和志愿者。可变翻转角(VFA)协议还适用于体模。$R_1$使用单体素拟合进行定量,并检查丘脑和小脑以及灰质和白质中区域的分布。排除边界拟合,并在整个浓度范围内量化弛豫率。通过Kolmogorov-Smirnov评分评估$R_1$分布的正态性,并通过双样本t检验评估序列间一致性。结果:IR法、SR法和VFA法的体模弛豫率分别为7.16 Hz/mM、9.22 Hz/mM和10.65 Hz/mM ~ 11.91 Hz/mM。体内$R_1$表现出较低的参与者内IR变异。SR $R_1$低于IR值,参与者间和参与者内的变异程度相同。用不同方案绘制的脑区和幻肢显著不同,t检验p值在0和5E-10之间。结论:结果表明,$R_1$和相应的弛豫率具有显著的方案依赖性,这表明应尝试进行方法间比较(如果有的话)。
英文摘要:
Purpose: Recent work highlights the breadth of reported spin-lattice relaxation rates ($R_1$) for individual tissues. One potential source of variation is the protocol used to determine $R_1$. The methodological dependence of R1 and relaxivity $r_1$ are investigated. Methods: $R_1$ is quantified in gel phantoms with varying concentration of MnCl2, and a small cohort of three healthy volunteers using different acquisition methods. Siemens inversion recovery (IR) and saturation recovery (SR) protocols are applied to phantoms and volunteers. Variable flip angle (VFA) protocols are additionally applied to phantoms. $R_1$ is quantified using single voxel fits, and distributions examined for regions in the thalamus, and cerebellum as well as grey and white matter. Phantoms exclude boundary fits and relaxivity is quantified across the full concentration range. Normality of $R_1$ distributions is assessed by Kolmogorov-Smirnov score, and inter-sequence agreement by two-sample t-test. Results: Phantom relaxivity is found to be 7.16 Hz/mM, 9.22 Hz/mM and 10.65 Hz/mM to 11.91 Hz/mM for IR, SR and VFA methods, respectively. In vivo $R_1$ exhibit low intra-participant variation for IR. SR $R_1$ are lower than IR values with inter- and intra-participant variation on the same order. Brain regions and phantoms mapped with different protocols varied significantly with t-test p-values between 0 and 5E-10. Conclusion: Results suggest a significant protocol dependence of $R_1$, and corresponding relaxivity, suggesting inter-method comparisons should be attempted tentatively, if at all.
机构、总结、中文摘要均由AI分析而来,可能存在错误,请以论文正文为准。