使用稀疏MP2RAGE校正磁化传递饱和(MTsat)图中的T1偏差
Correcting for T1 bias in Magnetization Transfer Saturation (MTsat) Maps Using Sparse-MP2RAGE
AI总结:
本研究比较VFA与csMP2RAGE T1成像在MTsat中的应用,证明csMP2RAGE可替代VFA,且扫描更快、可重复性高,并校正T1偏差。
摘要:
目的:磁化传递饱和(MTsat)成像常用于检查脑组织的大分子含量。本研究比较了可变翻转角(VFA)T1成像与压缩感知(cs)MP2RAGE T1成像在加速MTsat成像方面的表现。方法:在3特斯拉下,将VFA、MP2RAGE和csMP2RAGE与反转恢复(IR)T1在体模中进行比较。在四名健康受试者中采集相同的1毫米VFA、MP2RAGE和csMP2RAGE协议,以比较所得的T1和MTsat。使用Bloch-McConnell模拟来研究体模和体内T1结果之间的差异。最后,十名健康对照者使用csMP2RAGE MTsat协议进行了两次成像,以量化可重复性。结果:MP2RAGE和csMP2RAGE协议分别比VFA协议快13.7%和32.4%。所有方法在体模中均提供了准确的T1值(差异<5%),但VFA的T1时间准确性受T2差异的影响大于MP2RAGE。在体内,VFA产生的T1时间比MP2RAGE和csMP2RAGE更长。模拟表明,VFA与基于IR的方法(MP2RAGE和IR)之间T1值的偏差可能由反转脉冲的MT效应解释。在重测实验中,我们发现csMP2RAGE对T1成像的最小可检测变化为3%,对MTsat成像为7.9%。结论:我们证明了csMP2RAGE可以在MTsat协议中替代VFA T1成像。此外,使用csMP2RAGE序列可以实现更短的扫描时间和高可重复性。
英文摘要:
Purpose: Magnetization transfer saturation (MTsat) mapping is commonly used to examine the macromolecular content of brain tissue. This study compared variable flip angle (VFA) T1 mapping against compressed sensing (cs)MP2RAGE T1 mapping for accelerating MTsat imaging. Methods: VFA, MP2RAGE and csMP2RAGE were compared against inversion recovery (IR) T1 in a phantom at 3 Tesla. The same 1 mm VFA, MP2RAGE and csMP2RAGE protocols were acquired in four healthy subjects to compare the resulting T1 and MTsat. Bloch-McConnell simulations were used to investigate differences between the phantom and in vivo T1 results. Finally, ten healthy controls were imaged twice with the csMP2RAGE MTsat protocol to quantify repeatability. Results: The MP2RAGE and csMP2RAGE protocols were 13.7% and 32.4% faster than the VFA protocol, respectively. All approaches provided accurate T1 values (<5% difference) in the phantom, but the accuracy of the T1 times was more impacted by differences in T2 for VFA than for MP2RAGE. In vivo, VFA generated longer T1 times than MP2RAGE and csMP2RAGE. Simulations suggest that the bias in the T1 values between VFA and IR-based approaches (MP2RAGE and IR) could be explained by the MT-effects from the inversion pulse. In the test-retest experiment, we found that the csMP2RAGE has a minimum detectable change of 3% for T1 mapping and 7.9% for MTsat imaging. Conclusions: We demonstrated that csMP2RAGE can be used in place of VFA T1 mapping in an MTsat protocol. Furthermore, a shorter scan time and high repeatability can be achieved using the csMP2RAGE sequence.
机构、总结、中文摘要均由AI分析而来,可能存在错误,请以论文正文为准。