空气压陷法对圆锥角膜与正常角膜中央厚度的评价
Evaluation of central corneal thickness in keratoconus and normal corneas during air puff indentation
摘要:
目的:本研究旨在通过角膜可视化Scheimpflug技术(Corvis ST)研究圆锥角膜和正常角膜在空气抽吸压痕过程中中央角膜厚度(CCT)的实际变化。方法:对32例圆锥角膜患者和46例正常人进行了研究。选择CCT初始、CCT最终和CCT峰值三个参数分别代表空气抽吸压痕过程中初始时间、最终时间和最高角膜曲率时的CCT。采用Wilcoxon符号秩检验(配对样本检验)评估圆锥角膜组和正常组这3个参数之间的差异。进行单变量线性回归分析以确定CCT初始对CCT峰值和CCT最终的影响,以及每组中空气抽吸力对CCT的影响。用受试者工作特征(ROC)曲线评价3个参数的区分能力。结果如下:结果表明,圆锥角膜组和正常角膜组的CCT峰值和CCT最终值均较CCT初始值显著降低(P<0.01)。回归分析显示正常角膜组的CCT峰值与CCT初始值呈显著正相关(R2= 0.337,p <0.01),而圆锥角膜组的CCT峰值与CCT初始值无相关性(R2= 0.029,p =0.187)。同样地,吹气力和CCT的回归模型揭示了圆锥角膜和正常角膜组之间CCT变化的不同模式。ROC曲线显示CCT峰值的AUC最大,为0.940,准确性、敏感性和特异性分别为94.9%、87.5%和100%。结论:压缩空气压陷过程中角膜中央厚度发生变化,圆锥角膜与正常角膜中央厚度有显著差异。该变化规律有助于圆锥角膜的诊断,并为角膜生物力学研究奠定了基础。
英文摘要:
Purpose: This study aimed to investigate the actual changes of central corneal thickness (CCT) in keratoconus and normal corneas during air puff indentation, by using corneal visualization Scheimpflug technology (Corvis ST). Methods: A total of 32 keratoconic eyes and 46 normal eyes were included in this study. Three parameters of CCTinitial, CCTfinal and CCTpeak were selected to represent the CCT at initial time, final time and highest corneal concavity, respectively, during air puff indentation. Wilcoxon signed rank test (paired sample test) was used to assess the differences between these 3 parameters in both keratoconus and normal groups. Univariate linear regression analysis was performed to determine the effect of CCTinitial on CCTpeak and CCTfinal, as well as the impact of air puff force on CCT in each group. Receiver operating characteristic (ROC) curves were constructed to evaluate the discriminative ability of the 3 parameters. Results: The results demonstrated that CCTpeak and CCTfinal were significantly decreased (p<0.01) compared to CCTinitial in both keratoconus and normal groups. Regression analysis indicated a significant positive correlation between CCTpeak and CCTinitial in normal cornea group (R2=0.337, p<0.01), but not in keratoconus group (R2=0.029, p=0.187). Likewise, regression models of air puff force and CCT revealed the different patterns of CCT changes between keratoconus and normal cornea groups. Furthermore, ROC curves showed that CCTpeak exhibited the greatest AUC (area under ROC curve) of 0.940, with accuracy, sensitivity and specificity of 94.9%, 87.5% and 100%, respectively. Conclusions: CCT may change during air puff indentation, and is significantly different between keratoconus and normal cornea groups. The changing pattern is useful for the diagnosis of keratoconus, and lays the foundation for corneal biomechanics.
机构、总结、中文摘要均由AI分析而来,可能存在错误,请以论文正文为准。