不同氩基气体对MWPC探测器时间分辨率及特性的影响

Time resolution and characteristic study of MWPC detectors with different Argon based gas

Rajendra Nath Patra, R. N. Singaraju, S. Biswas, T. K. Nayak, Y. P. Viyogi

摘要:

制作了多丝正比室(MWPC)样机,研究其各种特性。该探测器包含镀金钨丝(20 $μm$直径)的阳极框架上,与2.8 mm的间距。阳极和阴极之间的间隙为3 mm,阳极和读出之间的间隙也为3 mm。详细的研究MWPC的增益,能量和定时分辨率和效率的测量已被执行。使用Ar/CO $_{2}$气体混合物以70:30和90:10的比例操作检测器。用该探测器获得了$^{55}$ Fe X射线源的能谱。利用X射线能谱计算了探测器的增益和能量分辨率。时间分辨率为$\sim$ 10 ns。

英文摘要:

A prototype of Multi-Wire Proportional Chambers (MWPC) has been fabricated for the study of its various characteristics. The detector contains gold-coated tungsten wires (20 $μm$ diameter) on the anode frame, with a pitch of 2.8 mm. The gap between the anode and the cathode is 3 mm and the gap between anode and read-out is also 3 mm. Detailed study of MWPC in terms of gain, energy and timing resolution and efficiency measurements have been performed. The detector has been operated using Ar/CO$_{2}$ gas mixtures with 70:30 and 90:10 ratio. Energy spectrum of $^{55}$Fe X-ray source is obtained for the detector. The gain and energy resolution of the detector were calculated using X-ray spectrum. Time resolution is obtained $\sim$10 ns.