arXivDaily arXiv每日学术速递 周一至周五更新
arXiv 1806.02494physics.ins-det
收藏

Nuclear Pulse Charge Measurement with a Method of Time over Linear Threshold

Zhengqi Song, Yonggang Wang, Yong Xiao, Qiang Cao

英文摘要:

Time over dynamic threshold (TODT) method, proposed in our previous work, has been successfully used for nuclear pulse charge measurement in PET detectors. It has advantages of strict linearity, large dynamic range, and better energy resolution, but requires a relatively complex circuit to generate specific dynamic threshold. In this paper, we propose to replace the dynamic threshold by a simpler linearly increasing threshold (called as TOLT method) with the aim to simplify the threshold generation circuit meanwhile maintaining its high energy resolution. Mathematical analysis on this replacement and the related realization circuit are presented. By energy spectrum measurement of PET detectors, the method is evaluated. The energy resolutions of PET detectors, composed of a PMT coupled with LYSO and LaBr3 crystal, are measured as 12.54% and 5.18% respectively, which is equivalent to the result obtained by TODT method. The test results show that the TOLT method is more practicable for charge measurement of nuclear detectors.

机构、总结、中文摘要均由AI分析而来,可能存在错误,请以论文正文为准。

↑