Bileaflet mechanical heart valves have one of the most successful valve designs for more than 30 years. These valves are often used for aortic valve replacement, where the geometry of the aortic root sinuses may vary due to valvular disease and affect valve performance. Common geometrical sinus changes may be due to valve stenosis and insufficiency. In the current study, the effect of these geometrical changes on the mean flow and velocity fluctuations downstream of the valve and aortic sinuses were investigated. The study focused on the fully-open leaflet position where blood velocities are close to their maximum. Simulation results were validated using previous experimental laser Doppler anemometry (LDA) measurements. Results showed that as the stenosis and insufficiency increased there were more flow separation and increased local mean velocity downstream of the leaflets. In addition, the detected elevated velocity fluctuations were associated with higher Reynolds shear stresses levels, which may increase the chances of blood damage and platelet activation and may lead to increased risk of blood clot formation.
Hemodynamics of a Bileaflet Mechanical Heart Valve with Different Levels of Dysfunction
Heart disease is one of leading causes of mortality worldwide. Healthy heart valves are key for proper heart function. When these valves dysfunction, a replacement is often necessary in severe cases. The current study presents an investigation of the pulsatile blood flow through a bileaflet mechanical heart valve (BMHV) where one leaflet is healthy and can fully open and the other leaflet cannot fully open with different levels of dysfunction. To better understand the implications that a dysfunctional leaflet has on the blood flow through these valves, analysis of flow characteristics such as …
Dynamic simulation of aortic valve stenosis using a lumped parameter cardiovascular system model with flow regime dependent valve pressure loss characteristics