超声组织谐波成像中重要参数变化对基于特征空间的最小方差权值计算的影响
Effects of Important Parameters Variations on Computing Eigenspace-Based Minimum Variance Weights for Ultrasound Tissue Harmonic Imaging
摘要:
近年来,最小方差(MV)波束形成由于其高分辨率和高对比度而在B超成像中得到了广泛的研究。然而,MV波束形成器的性能在噪声的存在下降低,作为不准确的协方差矩阵估计的结果,这导致低质量的图像。二次谐波成像(SHI)提供了许多优于常规脉冲回波USI的优点,例如增强的轴向和横向分辨率。然而,低信噪比(SNR)是SHI的主要问题。本文采用基于特征空间的最小方差(EIBMV)波束形成器进行二次谐波USI。组织谐波成像(THI)是通过脉冲反转(PI)技术实现的。使用EIBMV权重而不是MV权重将导致旁瓣减少和对比度提高,而不会损害MV波束形成器的高分辨率(即使在存在强噪声的情况下)。此外,我们还研究了计算EIBMV权重时重要参数变化的影响,即,K、L和δ对SHI中获得的分辨率和对比度的影响。用数值数据(点靶和囊肿模型)对结果进行了评价,并为THI指出了合适的EIBMV参数。
英文摘要:
In recent years, the minimum variance (MV) beamforming has been widely studied due to its high resolution and contrast in B-mode Ultrasound imaging (USI). However, the performance of the MV beamformer is degraded at the presence of noise, as a result of the inaccurate covariance matrix estimation which leads to a low quality image. Second harmonic imaging (SHI) provides many advantages over the conventional pulse-echo USI, such as enhanced axial and lateral resolutions. However, the low signal-to-noise ratio (SNR) is a major problem in SHI. In this paper, Eigenspace-based minimum variance (EIBMV) beamformer has been employed for second harmonic USI. The Tissue Harmonic Imaging (THI) is achieved by Pulse Inversion (PI) technique. Using the EIBMV weights, instead of the MV ones, would lead to reduced sidelobes and improved contrast, without compromising the high resolution of the MV beamformer (even at the presence of a strong noise). In addition, we have investigated the effects of variations of the important parameters in computing EIBMV weights, i.e., K, L, and δ, on the resolution and contrast obtained in SHI. The results are evaluated using numerical data (using point target and cyst phantoms), and the proper parameters of EIBMV are indicated for THI.
机构、总结、中文摘要均由AI分析而来,可能存在错误,请以论文正文为准。