EM重建多通道屏蔽式施源器用于宫颈癌近距离放射治疗的可行性研究

On EM Reconstruction of a Multi Channel Shielded Applicator for Cervical Cancer Brachytherapy: A Feasibility Study

D. Tho, E. Racine, H. Easton, W. Y. Song, L. Beaulieu

摘要:

电磁跟踪(EMT)是近距离放射治疗中自动化导管和施源器定位的一种有前途的技术。在这项工作中,提出了一个概念验证的屏蔽串联施源器专用于强度调制近距离放射治疗的各个通道的重建。重建了直原型的所有六个通道,并测量了两个相对通道之间的距离。研究了屏蔽体对EMT系统数据波动的影响。与CAD规定尺寸的差异小于2 mm。其中一个通道距离发生器更远的通道对具有更高的通道间距离和更高的可变性。在第一个110 cm重建中,所有通道间距离均在几何公差范围内。根据配对Student t检验,带和不带屏蔽施加器头端的EM系统给出的数据无显著差异。该研究表明,在施加器的机械精度内重建通道路径是可能的。

英文摘要:

Electromagnetic tracking (EMT) is a promising technology for automated catheter and applicator reconstruc- 10 tions in brachytherapy. In this work, a proof-of-concept is presented for reconstruction of the individual channels of a shielded tandem applicator dedicated to intensity modulated brachytherapy. All six channels of a straight prototype was reconstructed and the distance between two opposite channels was measured. A study was also conducted on the influence of the shield on the data fluctuation of the EMT system. The differences with the CAD specified dimensions are under 2 mm. The pair of channels which has one of it more distant from the generator have 15 higher inter-channel distance with higher variability. In the first 110 cm reconstruction, all inter-channel distances are within the geometrical tolerances. According to a paired Student t-test, the data given by the EM system with and without the shield applicator tip are not significantly different. This study shows that the reconstruction of channel path within the mechanical accuracy of the applicator is possible.