通过溅射钛表面的再污染测量监测溅射深度剖面分析设备的真空质量
Vacuum Quality Monitoring of Analytical Sputter Depth Profile Equipment by Recontamination Measurements of a Sputtered Ti Surface
摘要:
通常,表面敏感分析技术与通过惰性离子溅射的样品侵蚀相结合用于固态样品的成分深入分析。样品一层一层地被侵蚀,然后估计实际表面的成分。通过在轮廓测量的每个溅射深度中花费更多的时间进行测量,可以增加元素检测限。但在此期间,必须避免样品表面通过真空吸附而被研究元素再次污染。通常,分析仪器的真空质量是使用UHV系统的基准压力来监测的。本文提出了一种新的方法,它改进了这种粗糙的方法。溅射的Ti表面吸附的残留气体颗粒的再污染被用来监测溅射深度剖面特定的真空条件下的XPS微探针量子2000超过4年的时间。该方法适用于各种溅射深度剖面仪的状态监测。
英文摘要:
Often a surface sensitive analytical technique in combination with sample erosion by inert ion sputtering is used for compositional in-depth analysis of solid state samples. Layer by layer the sample gets eroded and then the composition of the actual surface is estimated. The elemental detection limits can be increased by spending more time for the measurement in each sputter depth of the profile measurement. But during this time a significant sample surface recontamination with the element under investigation via adsorption form the vacuum has to be avoided. Commonly the vacuum quality of an analytical instrument is monitored using the base pressure of the UHV system. This article presents a novel method, which improves this crude approach. The recontamination of a sputtered Ti surface by adsorbed residual gas particles was used to monitor the sputter depth profile specific vacuum condition of an XPS microprobe Quantum 2000 over a 4 years period. This new method is suitable to monitor the condition of every sputter depth profiling instrument.