灵活的A扫描速率MHz OCT:通过相干平均进行计算降尺度

Flexible A-scan rate MHz OCT: Computational downscaling by coherent averaging

Tom Pfeiffer, Wolfgang Wieser, Thomas Klein, Markus Petermann, Jan-Phillip Kolb, Matthias Eibl, Robert Huber

摘要:

为了实现线速可调的快速OCT系统,我们研究了基于FDML的MHz-OCT系统的图像数据平均。线速率可以在软件中降低,并以提高系统灵敏度和图像质量为代价。我们比较相干和非相干平均,以有效地按比例降低系统速度的3.2 MHz FDML OCT系统在后处理约100 kHz。我们表明,相干平均是可能的MHz系统,没有特殊的干涉仪设计或数字相位稳定。我们显示了在体内的人手指关节的OCT图像,具有非常高的质量和深穿透。

英文摘要:

In order to realize fast OCT-systems with adjustable line rate, we investigate averaging of image data from an FDML based MHz-OCT-system. The line rate can be reduced in software and traded in for increased system sensitivity and image quality. We compare coherent and incoherent averaging to effectively scale down the system speed of a 3.2 MHz FDML OCT system to around 100 kHz in postprocessing. We demonstrate that coherent averaging is possible with MHz systems without special interferometer designs or digital phase stabilisation. We show OCT images of a human finger knuckle joint in vivo with very high quality and deep penetration.