利用超顺磁性氧化铁纳米颗粒和交变磁场测量血液凝固的方法的开发

Development of a method for measuring blood coagulation using superparamagnetic iron oxide nanoparticles and an alternating magnetic field

Kenya Murase

摘要:

我们开发了一种使用超顺磁性氧化铁纳米颗粒(SPION)和交变磁场(AMF)测量血液凝固的方法。用梯度线圈检测AMF诱导的SPION与血液混合的三次和五次谐波信号。在不同温度和红细胞压积条件下,向绵羊全血中加入CaCl_2溶液,人工诱导血液凝固。我们通过将归一化的三次谐波信号的时间过程拟合为S(t)=(1-Sinf)exp(-kt)+Sinf来计算凝固率(k)和无限时间处的归一化信号强度(Sinf)。k值随温度升高而显著增加,随红细胞压积升高而显著降低。Sinf值随温度升高而显著降低,并随红细胞压积的增加而增加。通过向从小鼠采集的全血中加入肝素来诱导血液抗凝。在加入肝素后25 min或更长时间,加入和不加入肝素组之间的三次和五次谐波信号均存在显著差异。我们还计算了3次和5次谐波信号的粘度范围从0.001至1公斤/米/秒,假设自旋离子的磁化强度和粒度分布服从朗之万理论的顺磁性和对数正态分布,分别。第三和第五谐波信号随着粘度的增加而缓慢增加,并且分别在约0.015和0.025 kg/m/s处具有峰值。在这些峰值之后,它们随着粘度的增加而单调下降。这些结果证实了我们的方法的基本原理。总之,我们的方法将是有用的测量血液凝固和抗凝血,并为研究他们的过程。

英文摘要:

We developed a method for measuring blood coagulation using superparamagnetic iron oxide nanoparticles (SPIONs) and an alternating magnetic field (AMF). The 3rd and 5th harmonic signals from SPIONs mixed with blood induced by AMF were detected using a gradiometer coil. Blood coagulation was induced artificially by adding CaCl2 solution to whole blood of sheep at various temperatures and hematocrits. We calculated the coagulation rate (k) and normalized signal intensity at infinite time (Sinf) by fitting the time course of the normalized 3rd harmonic signal to S(t)=(1-Sinf)exp(-kt)+Sinf. The k values increased significantly with increasing temperature and decreased significantly with increasing hematocrit. The Sinf values decreased significantly with increasing temperature and tended to increase with increasing hematocrit. Blood anticoagulation was induced by adding heparin to the whole blood sampled from mice. There were significant differences in both the 3rd and 5th harmonic signals between groups with and without heparin at 25 min or more after adding heparin. We also calculated the 3rd and 5th harmonic signals for viscosities ranging from 0.001 to 1 kg/m/s, with an assumption that the magnetization and particle size distribution of SPIONs obey the Langevin theory of paramagnetism and log-normal distribution, respectively. The 3rd and 5th harmonic signals increased slowly with increasing viscosity and had peaks at approximately 0.015 and 0.025 kg/m/s, respectively. After these peaks, they decreased monotonically with increasing viscosity. These results confirm the rationale of our method. In conclusion, our method will be useful for measuring blood coagulation and anticoagulation and for studying their processes.

机构、总结、中文摘要均由AI分析而来,可能存在错误,请以论文正文为准。

↑