DICE校准项目:设计、表征和初步结果

The DICE calibration project: design, characterization, and first results

N. Regnault, A. Guyonnet, K. Schahmanèche, L. Le Guillou, P. Antilogus, P. Astier, E. Barrelet, M. Betoule, S. Bongard, J. -C. Cuillandre, C. Juramy, R. Pain, P… 展开作者

N. Regnault, A. Guyonnet, K. Schahmanèche, L. Le Guillou, P. Antilogus, P. Astier, E. Barrelet, M. Betoule, S. Bongard, J. -C. Cuillandre, C. Juramy, R. Pain, P. -F. Rocci, P. Tisserand, F. Villa

摘要:

我们描述的设计,操作,和第一个光度校准项目,称为DICE(直接照明校准实验),旨在实现精确的仪器校准的光学望远镜的结果。DICE的核心是一个照明设备,由24个窄光谱、高强度的发光二极管(LED)组成,选择这些发光二极管覆盖紫外线到近红外光谱范围。它实现了一个点状源放置在一个有限的距离,从望远镜的入射光瞳,产生一个平场照明,覆盖整个视场的成像器。该系统的目的是执行一个轻量级的例行监测的成像仪通带的精度优于5 per-mil的相对通带归一化和约3 μ m的滤波器截止位置。在分光光度计工作台上校准光源。作为我们的基本计量标准,我们使用在NIST校准的光电二极管。绘制每个光束的辐射强度,并测量每个LED的光谱。所有测量均在0°C至25°C的温度范围内进行,以研究系统的温度依赖性。将光度测量和光谱测量结合到一个模型中,该模型预测源的光谱强度作为温度的函数。我们发现,校准光束是稳定的$10^{-4}$水平-考虑到轻微的温度依赖性的LED发光特性后。我们表明,源的光谱强度的特点是可以与波长的精度为30 μ m。在通量中,我们达到约0.2-0.5%的精度,这取决于我们如何理解影响NIST光电二极管校准的误差预算的非对角项。通过常规的60-mn校准程序,该装置能够将通带限制在目标精度水平。

英文摘要:

We describe the design, operation, and first results of a photometric calibration project, called DICE (Direct Illumination Calibration Experiment), aiming at achieving precise instrumental calibration of optical telescopes. The heart of DICE is an illumination device composed of 24 narrow-spectrum, high-intensity, light-emitting diodes (LED) chosen to cover the ultraviolet-to-near-infrared spectral range. It implements a point-like source placed at a finite distance from the telescope entrance pupil, yielding a flat field illumination that covers the entire field of view of the imager. The purpose of this system is to perform a lightweight routine monitoring of the imager passbands with a precision better than 5 per-mil on the relative passband normalisations and about 3Å on the filter cutoff positions. The light source is calibrated on a spectrophotometric bench. As our fundamental metrology standard, we use a photodiode calibrated at NIST. The radiant intensity of each beam is mapped, and spectra are measured for each LED. All measurements are conducted at temperatures ranging from 0°C to 25°C in order to study the temperature dependence of the system. The photometric and spectroscopic measurements are combined into a model that predicts the spectral intensity of the source as a function of temperature. We find that the calibration beams are stable at the $10^{-4}$ level -- after taking the slight temperature dependence of the LED emission properties into account. We show that the spectral intensity of the source can be characterised with a precision of 3Å in wavelength. In flux, we reach an accuracy of about 0.2-0.5% depending on how we understand the off-diagonal terms of the error budget affecting the calibration of the NIST photodiode. With a routine 60-mn calibration program, the apparatus is able to constrain the passbands at the targeted precision levels.

机构、总结、中文摘要均由AI分析而来,可能存在错误,请以论文正文为准。

↑