一种伪消色差中红外Mueller椭偏仪
A mid-infrared Mueller ellipsometer with pseudo-achromatic optical elements
摘要:
本文的目的是提出一种新的宽带米勒椭偏仪设计工作在中红外范围内,从3至14微米。Mueller椭偏仪可安装在反射或透射配置中,由偏振态发生器(PSG)、样品架和偏振态分析器(PSA)组成。PSG由一个线性偏振器和一个延迟器组成,它们依次旋转以产生一组四个最佳偏振态。延迟器包括由两个相同的菲涅耳菱形对称设置并通过光学接触连接而制成的双棱镜,从而使整体呈“V”形。延迟是由光束在通过双棱镜传播时经历的四次全内反射引起的。全内反射允许产生准消色差延迟。PSA与PSG相同,但其光学元件的安装顺序相反。在测量运行后,仪器产生一组16个独立值,这是计算样品的Mueller矩阵所需的最小数据量。Mueller椭偏仪的设计基于客观标准的优化,该标准允许最小化从原始数据到样品的Mueller矩阵的误差传播。伪消色差光学元件确保所有波长的测量质量均匀。米勒椭偏仪的精度和准确度方面的性能,进行了讨论,并说明了几个例子。
英文摘要:
The purpose of this article is to present a new broadband Mueller ellipsometer designed to work in the mid-infrared range, from 3 to 14 microns. The Mueller ellipsometer, which can be mounted in reflection or in transmission configuration, consists of a polarization state generator (PSG), a sample holder, and a polarization state analyzer (PSA). The PSG consists in one linear polarizer and a retarder sequentially rotated to generate a set of four optimal polarization states. The retarder consists in a bi-prism made of two identical Fresnel rhombs disposed symmetrically and joined by optical contact, giving the ensemble a "V" shape. Retardation is induced by the four total internal reflections that the beam undergoes when it propagates through the bi-prism. Total internal reflection allows to generate a quasi-achromatic retardation. The PSA is identical to the PSG, but with its optical elements mounted in reverse order. After a measurement run, the instrument yields a set of sixteen independent values, which is the minimum amount of data required to calculate the Mueller matrix of the sample. The design of the Mueller ellipsometer is based on the optimization of an objective criterion that allows minimizing the propagation of errors from raw data to the Mueller matrix of the sample. The pseudo-achromatic optical elements ensure a homogeneous quality of the measurements for all wavelengths. The performance of the Mueller ellipsometer in terms of precision, and accuracy, is discussed and illustrated with a few examples.
机构、总结、中文摘要均由AI分析而来,可能存在错误,请以论文正文为准。