快速近红外Mueller矩阵椭偏仪的动态响应
Dynamic Response of a fast near infra-red Mueller matrix ellipsometer
摘要:
研究了近红外铁电液晶基Mueller矩阵椭偏仪(NIR FLC-MME)的动态响应特性。一个时间相关的仿真模型,使用测量的时间响应的各个FLC,被用来描述测量的时间响应。此外,检测器和前置放大器的脉冲响应的特征在于,并包括在仿真模型中。在模拟中再现了MME的测量的时间相关强度响应,并且得出结论,FLC的切换时间是基于FLC的MME的Mueller矩阵测量时间的限制因素。基于测量和模拟,我们的基于FLC的NIR-MME系统估计以每次Mueller矩阵测量约16 ms的最大速度操作。FLC-MME的操作速度可以快几倍,因为晶体的切换时间取决于被切换的单个晶体以及它被切换到什么状态。作为演示,展示了测量的Mueller矩阵的时间响应和厚液晶可变延迟器在改变状态时的延迟。
英文摘要:
The dynamic response of a near infrared Ferroelectric Liquid Crystal based Mueller matrix ellipsometer (NIR FLC-MME) is presented. A time dependent simulation model, using the measured time response of the individual FLCs, is used to describe the measured temporal response. Furthermore, the impulse response of the detector and the pre-amplifier is characterized and included in the simulation model. The measured time-dependent intensity response of the MME is reproduced in simulations, and it is concluded that the switching time of the FLCs is the limiting factor for the Mueller matrix measurement time of the FLC-based MME. Based on measurements and simulations our FLC based NIR-MME system is estimated to operate at the maximum speed of approximately 16 ms per Mueller matrix measurement. The FLC-MME may be operated several times faster, since the switching time of the crystals depends on the individual crystal being switched, and to what state it is switched. As a demonstration, the measured temporal response of the Mueller matrix and the retardance of a thick liquid crystal variable retarder upon changing state is demonstrated.
机构、总结、中文摘要均由AI分析而来,可能存在错误,请以论文正文为准。