水中177 MeV质子束的核晕:理论、测量和参数化

Nuclear halo of a 177 MeV proton beam in water: theory, measurement and parameterization

Bernard Gottschalk, Ethan W. Cascio, Juliane Daartz, Miles S. Wagner

摘要:

单能笔形束流在水中的剂量分布由一个电磁“核”、一个荷电核秒的“晕”和一个更大的中性秒的“光环”组成。这些区域重叠,但每个区域都有不同的空间特征。我们已经测量的核心/晕使用177 MeV的测试束偏移在水箱中。射束监测器是注量校准的平面平行电离室(IC)和场室,剂量校准Exradin T1,因此剂量测量是绝对的(MeV/g/p)。我们进行了深度剂量扫描,在10个位移从束轴范围从0到10厘米。剂量跨越五个数量级,从晕到光环的转变清晰可见。我们对数据进行了模型依赖(MD)和模型独立(MI)拟合。MD拟合将剂量分为核心、弹性/非弹性核、非弹性核和先兆项,并实现15%的全局rms测量/拟合比。MI拟合使用三次样条,相同的比率为9%。我们回顾文献,特别是使用佩德罗尼的参数化的核心/晕。几篇论文改进了他的晕的高斯横向分布,但都保留了他的T(w),深度剂量的径向积分乘以核心和晕项,并激励测量与大型“布拉格峰室”(BPC)。我们认为,这种使用T(w),它的定义包括能量沉积的核秒数,是不正确的。T(w)在核心项中,至少在晕的一部分中,应该用纯电磁质量阻止本领来代替。BPC测量是不必要的,并且与笔形波束的参数化无关。

英文摘要:

The dose distribution of a monoenergetic pencil beam in water consists of an electromagnetic "core", a "halo" from charged nuclear secondaries, and a much larger "aura" from neutral secondaries. These regions overlap, but each has distinct spatial characteristics. We have measured the core/halo using a 177MeV test beam offset in a water tank. The beam monitor was a fluence calibrated plane parallel ionization chamber (IC) and the field chamber, a dose calibrated Exradin T1, so the dose measurements are absolute (MeV/g/p). We performed depth-dose scans at ten displacements from the beam axis ranging from 0 to 10cm. The dose spans five orders of magnitude, and the transition from halo to aura is clearly visible. We have performed model-dependent (MD) and model-independent (MI) fits to the data. The MD fit separates the dose into core, elastic/inelastic nuclear, nonelastic nuclear and aura terms, and achieves a global rms measurement/fit ratio of 15%. The MI fit uses cubic splines and the same ratio is 9%. We review the literature, in particular the use of Pedroni's parametrization of the core/halo. Several papers improve on his Gaussian transverse distribution of the halo, but all retain his T(w), the radial integral of the depth-dose multiplying both the core and halo terms and motivating measurements with large "Bragg peak chambers" (BPCs). We argue that this use of T(w), which by its definition includes energy deposition by nuclear secondaries, is incorrect. T(w) should be replaced in the core term, and in at least part of the halo, by a purely electromagnetic mass stopping power. BPC measurements are unnecessary, and irrelevant to parameterizing the pencil beam.

机构、总结、中文摘要均由AI分析而来,可能存在错误,请以论文正文为准。

↑