S. Afach, G. Bison, K. Bodek, F. Burri, Z. Chowdhuri, M. Daum, M. Fertl, B. Franke, Z. Grujic, V. Helaine, R. Henneck, M. Kasprzak, K. Kirch, H. -C. Koch, A. Kozela, J. Krempel, B. Lauss, T. Lefort, Y. Lemiere, M. Meier, O. Naviliat-Cuncic, F. M. Piegsa, G. Pignol, C. Plonka-Spehr, P. N. Prashanth, G. Quemener, D. Rebreyend, S. Roccia, P. Schmidt-Wellenburg, A. Schnabel, N. Severijns, J. Voigt, A. Weis, G. Wyszynski, J. Zejma, J. Zenner, G. Zsigmond

摘要:

在这项工作中所描述的周围场补偿(SFC)系统安装在周围的四层金属磁屏蔽的中子电偶极矩谱仪位于保罗谢勒研究所。SFC系统将外部磁场的DC分量减小约20倍。在约2.5 m x 2.5 m x 3 m的控制体积内,磁场的干扰在$10^{-3}$ Hz至0.5 Hz的带宽下衰减5至50倍,这对应于长于数百秒的积分时间,代表了nEDM测量的重要时间尺度。这些屏蔽因子适用于来自任意源的随机环境噪声。这是通过比例积分反馈稳定系统来实现的,该比例积分反馈稳定系统包括比例因子的正则化伪逆矩阵,该比例因子将所有传感器位置处的磁场变化与SFC线圈中的电流变化相关联。

英文摘要:

The Surrounding Field Compensation (SFC) system described in this work is installed around the four-layer Mu-metal magnetic shield of the neutron electric dipole moment spectrometer located at the Paul Scherrer Institute. The SFC system reduces the DC component of the external magnetic field by a factor of about 20. Within a control volume of approximately 2.5m x 2.5m x 3m disturbances of the magnetic field are attenuated by factors of 5 to 50 at a bandwidth from $10^{-3}$ Hz up to 0.5 Hz, which corresponds to integration times longer than several hundreds of seconds and represent the important timescale for the nEDM measurement. These shielding factors apply to random environmental noise from arbitrary sources. This is achieved via a proportional-integral feedback stabilization system that includes a regularized pseudoinverse matrix of proportionality factors which correlates magnetic field changes at all sensor positions to current changes in the SFC coils.