Attenuation Correction of L-shell X-ray Fluorescence Computed Tomography Imaging

Liu Long, Huang Yang, Ma Bo, Xu Qing, Yan Lingtong, Li Li, Feng Songlin, Feng Xiangqian

摘要:

X射线荧光计算机断层成像(XFCT)是一种用于研究样品中元素空间分布的常用实验技术。某些元素的L壳XFCT灵敏度低于K壳XFCT。然而,该技术的图像重建比透射层析成像有更多的问题,其中之一是自吸收。在目前的工作中,开发了一种新的策略来处理这些问题。但对于L-shell XFCT的衰减校正研究较少,而衰减校正是获得精确重建图像的关键。我们利用已知的量和未知的感兴趣的元素浓度来表示未知的衰减图。在MLEM重建方法中,将衰减图加入到像素的贡献值中。结果表明,该方法的相对误差小于14.1%,能够很好地校正L壳XFCT。

英文摘要:

X-ray Fluorescence Computed Tomography(XFCT) is a prevalent experimental technique which is utilized to investigate the spatial distribution of elements in sample. The sensitivity of L-shell XFCT of some elements is lower than that of K-shell XFCT. However, the image reconstruction for this technique has much more problems than that of transmission tomography, one of which is self-absorption. In the present work, a novel strategy was developed to deal with such problems. But few researches are concerned on attenuation correction of L-shell XFCT that are essential to get accurate reconstructed image. We make use of the known quantities and the unknown elemental concentration of interest to express the unknown attenuation maps. And then the attenuation maps are added in the contribution value of the pixel in MLEM reconstruction method. Results indicate that the relative error is less than 14.1%, which is proved this method can correct L-shell XFCT very well.