监测Varian 21 EX和TrueBeam直线加速器的反向散射因子:测量和Monte Carlo建模
Monitor backscatter factors for the Varian 21EX and TrueBeam linear accelerators: measurements and Monte Carlo modeling
摘要:
目的:在放射治疗剂量计算中,必须考虑进入直线加速器监测室的背散射辐射(BSR)。在蒙特卡罗(MC)计算中,BSR可以明确建模,但仅当治疗头几何形状可用时。在这项研究中,监控器的后向散射因子(MBSFs),定义为一个参考字段的监控室中收集的电荷的比例,一个给定的字段,已被实验评估,并纳入MC建模的直线加速器与已知或未知的治疗头的几何形状。材料与方法:采用与Kubo(1989)相似的技术。然而,代替铅狭缝,使用1.8 mm直径的准直器和位于扩展SDD处的小型(2 mm直径)探测器。该设置为源提供了小于3.1 mm的视场,并允许从1 x 1到40 x 40 cm 2的开放视场的MBSF测量。在MC计算中,使用瓦里安21 EX直线加速器的6 MV和18 MV束流的BEAMnrc和DOSXYZnrc代码也明确地模拟了MBSF。证明了用于推导MC绝对剂量计算中使用的Dchforward值的方法。这些值来自两个21 EX和四个TrueBeam能量的测量MBSF。结果:测量了Varian 21 EX的6 MV和18 MV束流以及Varian TrueBeam直线加速器的6 MV、10 MV-FFF、10 MV和15 MV束流的MBSF。对于在本研究中为21 EX建模的开放场尺寸,测量的MBSF与MC计算值在统计(0.4%)和实验(0.2%)不确定性范围内一致。与21 EX Varian直线加速器相比,TrueBeam的MBSF在场尺寸上的变化大约小两倍。结论:测量的MBSFs和推导出的Dchforward因子允许将BSR效应纳入精确的放射治疗剂量计算中,而无需明确的后向散射建模。
英文摘要:
Purpose/Objective: Backscattered radiation (BSR) into linac monitor chamber has to be accounted for in radiotherapy dose calculations. In Monte Carlo (MC) calculations, the BSR can be modeled explicitly, but only when treatment head geometry is available. In this study, monitor backscatter factors (MBSFs), defined as the ratio of the charge collected in the monitor chamber for a reference field to that of a given field, have been evaluated experimentally and incorporated into MC modelling of linacs with either known or unknown treatment head geometry. Materials and methods: A telescopic technique similar to that by Kubo (1989) was used. However, instead of lead slits, a 1.8 mm diameter collimator and a small (2 mm diameter) detector positioned at extended SDD were used. This setup provided a field of view to the source of less than 3.1 mm and allowed for MBSF measurements of open fields from 1 x 1 to 40 x 40 cm2. MBSFs were also explicitly modeled in MC calculations using BEAMnrc and DOSXYZnrc codes for 6MV and 18MV beams of a Varian 21EX linac. A method for deriving the Dchforward values that are used in MC absolute dose calculations was demonstrated. These values were derived from measured MBSFs for two 21EX and four TrueBeam energies. Results: MBSFs were measured for 6MV and 18MV beams from Varian 21EX, and for 6MV, 10MV-FFF, 10MV, and 15MV beams from Varian TrueBeam linacs. For the open field sizes modeled in this study for the 21EX, the measured MBSFs agreed with MC calculated values within combined statistical (0.4%) and experimental (0.2%) uncertainties. Variation of MBSFs across field sizes was about a factor of two smaller for the TrueBeam compared to 21EX Varian linacs. Conclusions: Measured MBSFs and the derived Dchforward factors allow for the incorporation of the BSR effect into accurate radiotherapy dose calculations without explicit backscatter modelling.
机构、总结、中文摘要均由AI分析而来,可能存在错误,请以论文正文为准。