离子束中的吸收剂量:基于电离和基于注量测量的比较
Absorbed Dose in Ion Beams: Comparison of Ionisation and Fluence-based Measurements
摘要:
本文介绍了荧光核径迹探测器(FNTDs)和套管电离室的直接比较测量。使用单能质子(142.66 MeV,3*10^6 1/cm 2)和碳离子(270.55 MeV/u,3*10^6 1/cm 2)进行辐照。结果发现,如果通过有效阻止功率来解释较慢的质子和氦秒数,则在质子的情况下,通过使用FNTD的注量测量确定的水吸收剂量值与电离室测量具有良好的一致性(2.4%)。然而,对于碳,观察到4.5%的显著差异,这不能用碎片化、不确定性或实验设计来解释。结果表明W值为32.10 eV +/-2.6%。此外,没有观察到从Monte-Carlo输运模拟中预期的次级质子丰度。
英文摘要:
A direct comparison measurement of fluorescent nuclear track detectors (FNTDs) and a thimble ionisation chamber is presented. Irradiations were performed using monoenergetic protons (142.66 MeV, 3*10^6 1/cm2) and carbon ions (270.55 MeV/u, 3*10^6 1/cm2). It was found that absorbed dose to water values as determined by fluence measurements using FNTDs are, in case of protons, in good agreement (2.4 %) with ionization chamber measurements, if slower protons and Helium secondaries were accounted for by an effective stopping power. For carbon, however, a significant discrepancy of 4.5 % was seen, which could not be explained by fragmentation, uncertainties or experimental design. The results rather suggest a W-value of 32.10 eV +/- 2.6 %. Additionally, the abundance of secondary protons expected from Monte-Carlo transport simulation was not observed.
机构、总结、中文摘要均由AI分析而来,可能存在错误,请以论文正文为准。