SCUBA-2:使用亚毫米标准源的空中校准

SCUBA-2: on-sky calibration using submillimetre standard sources

Jessica T. Dempsey, Per Friberg, Tim Jenness, Remo P. J. Tilanus, Holly S. Thomas, Wayne S. Holland, Dan Bintley, David S. Berry, Edward L. Chapin, Antonio Chry… 展开作者

Jessica T. Dempsey, Per Friberg, Tim Jenness, Remo P. J. Tilanus, Holly S. Thomas, Wayne S. Holland, Dan Bintley, David S. Berry, Edward L. Chapin, Antonio Chrysostomou, Gary R. Davis, Andrew G. Gibb, Harriet Parsons, E. Ian Robson

摘要:

SCUBA-2是詹姆斯·克拉克·麦克斯韦望远镜(James Clerk Maxwell Telescope,JCMT)上的一台10000辐射热计亚毫米相机。仪器调试于2011年9月完成,全面科学运行于2011年10月开始。为了充分利用这一强大的新天文工具的潜力,仪器校准必须准确且易于理解。为此,改进了视线不透明度的计算方法,并给出了SCUBA-2仪器两个波段的大气消光关系。来自500多个初级和次级校准器观测的结果允许精确确定850和450微米阵列的通量转换因子(FCF)。介绍了仪器的光束形状和测光方法。校准因子确定良好,在850微米处的相对校准精度优于5%,在450微米处的相对校准精度优于10%,反映了导出的不透明度关系的成功以及仪器性能在几个月内的稳定性。校准观测的样本量和精确的FCF允许确定几个著名的亚毫米源的850和450微米通量,并将这些结果与SCUBA以前的测量结果进行比较。

英文摘要:

SCUBA-2 is a 10000-bolometer submillimetre camera on the James Clerk Maxwell Telescope (JCMT). The instrument commissioning was completed in September 2011, and full science operations began in October 2011. To harness the full potential of this powerful new astronomical tool, the instrument calibration must be accurate and well understood. To this end, the algorithms for calculating the line-of-sight opacity have been improved, and the derived atmospheric extinction relationships at both wavebands of the SCUBA-2 instrument are presented. The results from over 500 primary and secondary calibrator observations have allowed accurate determination of the flux conversion factors (FCF) for the 850 and 450 micron arrays. Descriptions of the instrument beam-shape and photometry methods are presented. The calibration factors are well determined, with relative calibration accuracy better than 5 per cent at 850 microns and 10 per cent at 450 microns, reflecting the success of the derived opacity relations as well as the stability of the performance of the instrument over several months. The sample-size of the calibration observations and accurate FCFs have allowed the determination of the 850 and 450 micron fluxes of several well-known submillimetre sources, and these results are compared with previous measurements from SCUBA.

机构、总结、中文摘要均由AI分析而来,可能存在错误,请以论文正文为准。

↑