用于大输入电容电导测量的宽带电流源
Wide-band current preamplifier for conductance measurements with large input capacitance
摘要:
提出了一种基于复合运算放大器的宽带电流放大器。结果表明,通过提高复合放大器的有效开环增益,可以显著提高放大器的带宽。所述电流增益为10 $^7$ V/A的放大器显示带宽约为100 kHz,输入并联电容为1 nF。放大器的电流噪声在1 kHz时测得约为46 fA/ $\sqrt{\rm Hz}$,接近设计噪声最小值。发现在1 kHz时的电压噪声约为2.9 nV/ $\sqrt{\rm Hz}$,这与输入级中使用的运算放大器的预期值非常一致。通过分析总噪声产生的频率范围,我们发现适合于快速锁定测量的最佳频率范围是从1 kHz到2 kHz。为了获得相同的信噪比,所报道的积分时间大约是传统积分时间的10%。
英文摘要:
A wide-band current preamplifier based on a composite operational amplifier is proposed. It has been shown that the bandwidth of the preamplifier can be significantly increased by enhancing the effective open-loop gain of the composite preamplifier. The described preamplifier with current gain 10$^7$ V/A showed the bandwidth of about 100 kHz with 1 nF input shunt capacitance. The current noise of the amplifier was measured to be about 46 fA/$\sqrt{\rm Hz}$ at 1 kHz, close to the design noise minimum. The voltage noise was found to be about 2.9 nV/$\sqrt{\rm Hz}$ at 1 kHz, which is in a good agreement with the value expected for the operational amplifier used in the input stage. By analysing the total noise produced by the preamplifier we found the optimal frequency range suitable for the fast lock-in measurements to be from 1 kHz to 2 kHz. To get the same signal-to-noise ratio, the reported preamplifier requires roughly 10% of the integration time used in measurements made with a conventional preamplifier.