摘要:
这项工作的目的是模拟髋关节假体在机械冲击下的行为。在人工髋关节置换术中,在下肢摆动阶段,假体头与髋臼杯之间可能发生微小分离。研究了两种不同尺寸的股骨头:直径28 mm和32 mm,分别由氧化铝和氧化锆制成。使用有限元分析(FEA),Abaqus软件数值确定冲击诱导应力。研究了倾角、力、材料和微分离的影响。此外,还从概率模型Todinov方法中开发了一种算法来预测股骨头和髋臼杯的寿命。模拟结果表明,最大拉伸应力达到杯的边缘附近的表面上。最差情况是安装在30^{ \circ }处的杯头。所有的模拟和测试表明,散装氧化锆比散装氧化铝具有更大的抗冲击性。当材料中存在大于0.7%体积的孔隙率时,失效概率可能大于0.9。模拟结果与实验结果吻合较好。这些试验和模拟结果对陶瓷修复体的寿命预测具有一定的指导意义。
英文摘要:
The aim of this work was to simulate the behaviour of hip prostheses under mechanical shocks. When hip joint is replaced by prosthesis, during the swing phase of the leg, a microseparation between the prosthetic head and the cup could occur. Two different sizes of femoral heads were studied: 28 and 32 mm diameter, made, respectively, in alumina and zirconia. The shock-induced stress was determined numerically using finite element analysis (FEA), Abaqus software. The influence of inclination, force, material, and microseparation was studied. In addition, an algorithm was developed from a probabilistic model, Todinov's approach, to predict lifetime of head and cup. Simulations showed maximum tensile stresses were reached on the cup's surfaces near to rim. The worst case was the cup-head mounted at 30^{\circ}. All simulations and tests showed bulk zirconia had a greater resistance to shocks than bulk alumina. The probability of failure could be bigger than 0.9 when a porosity greater than 0.7% vol. is present in the material. Simulating results showed good agreement with experimental results. The tests and simulations are promising for predicting the lifetime of ceramic prostheses.
机构、总结、中文摘要均由AI分析而来,可能存在错误,请以论文正文为准。