用热交换器进行氙再循环净化

Xenon Recirculation-Purification with a Heat Exchanger

K. L. Giboni, E. Aprile, B. Choi, T. Haruyama, R. F. Lang, K. E. Lim, A. J. Melgarejo, G. Plante

摘要:

基于液体氙的粒子探测器在过去几年中在尺寸上已经显著地增长,并且现在超过了一吨级。所需的高纯度氙气通常是通过高温吸气剂连续再循环氙气来实现的。这对冷却这些大型探测器的传统方式提出了挑战,因为在隔热良好的探测器中,大部分冷却功率都用于补偿再循环的损失。再冷凝过程中的相变需要比将气体从环境温度冷却到-100 ℃(173 K)多五倍的冷却功率。因此,为了降低大型探测器的冷却功率要求,我们建议通过热交换器使用来自净化的进入气体的热量来蒸发排出的氙气。一般来说,热交换器似乎只有非常有限的用途,因为蒸发和液化发生在零温差。然而,再循环泵的使用降低了提取液体的压力,迫使其蒸发,从而使其冷却。我们表明,这种温差可以用于一个有效的热交换过程。我们调查使用一个商业平行板换热器与小型液体氙探测器。虽然我们预期将受到可用冷却功率的限制,流量为约2 SLPM,但超过12 SLPM的流量可以很容易地维持,仅受泵速和流动回路阻抗的限制。热交换器以(96.8 +/- 0.5)%的效率运行。这为大规模实验中快速氙气再循环提供了可能性,同时最大限度地减少了热损失。

英文摘要:

Liquid-xenon based particle detectors have been dramatically growing in size during the last years, and are now exceeding the one-ton scale. The required high xenon purity is usually achieved by continuous recirculation of xenon gas through a high-temperature getter. This challenges the traditional way of cooling these large detectors, since in a thermally well insulated detector, most of the cooling power is spent to compensate losses from recirculation. The phase change during recondensing requires five times more cooling power than cooling the gas from ambient temperature to -100C (173 K). Thus, to reduce the cooling power requirements for large detectors, we propose to use the heat from the purified incoming gas to evaporate the outgoing xenon gas, by means of a heat exchanger. Generally, a heat exchanger would appear to be only of very limited use, since evaporation and liquefaction occur at zero temperature difference. However, the use of a recirculation pump reduces the pressure of the extracted liquid, forces it to evaporate, and thus cools it down. We show that this temperature difference can be used for an efficient heat exchange process. We investigate the use of a commercial parallel plate heat exchanger with a small liquid xenon detector. Although we expected to be limited by the available cooling power to flow rates of about 2 SLPM, rates in excess of 12 SLPM can easily be sustained, limited only by the pump speed and the impedance of the flow loop. The heat exchanger operates with an efficiency of (96.8 +/- 0.5)%. This opens the possibility for fast xenon gas recirculation in large-scale experiments, while minimizing thermal losses.

机构、总结、中文摘要均由AI分析而来,可能存在错误,请以论文正文为准。

↑