摘要:
HgCdTe近红外探测器中的通量相关非线性(互易性故障)会严重影响仪器的性能,特别是在精密光度测量方面。这种影响的原因目前还不清楚。为了研究互易性失效,建立了专用的测试系统。对于1和50,000光子/s之间的通量水平,实现了约0.1%/十倍的互易性故障的灵敏度。在1.7微米HgCdTe探测器中测量到约0.35%/十倍的互惠失效导致的与波长无关的非线性。
英文摘要:
Flux dependent non-linearity (reciprocity failure) in HgCdTe near infrared detectors can severely impact an instrument's performance, in particular with respect to precision photometric measurements. The cause of this effect is presently not understood. To investigate reciprocity failure, a dedicated test system was built. For flux levels between 1 and 50,000 photons/s, a sensitivity to reciprocity failure of approximately 0.1%/decade was achieved. A wavelength independent non-linearity due to reciprocity failure of about 0.35%/decade was measured in a 1.7 micron HgCdTe detector.