用数字符合计数、液体闪烁计数和限定立体角计数法标定241 Am
Standardization of 241Am by digital coincidence counting, liquid scintillation counting and defined solid angle counting
摘要:
核素241 Am通过α发射衰变为237 Np。大部分衰变(84.6%)位于能量为59.54 keV的237 Np激发态。应用数字符合计数法对241 Am溶液进行了标准化,方法是采用α-γ符合计数法和效率外推法。电子歧视与加压的正比计数器和结果进行了比较,与其他两个独立的技术:液体闪烁计数使用的逻辑和双符合在TDCR阵列和定义的立体角计数考虑到活性沉积物中的活性不均匀性。结果表明,三种方法之间的一致性在0.3%的限度内。2009年期间,将向国际参考系(SIR)发送一份该溶液的样本。对这三种方法的误差进行了详细的分析和比较。
英文摘要:
The nuclide 241Am decays by alpha emission to 237Np. Most of the decays (84.6 %) populate the excited level of 237Np with energy of 59.54 keV. Digital Coincidence Counting was applied to standardize a solution of 241Am by alpha-gamma coincidence counting with efficiency extrapolation. Electronic discrimination was implemented with a pressurized proportional counter and the results were compared with two other independent techniques: Liquid Scintillation Counting using the logical sum of double coincidences in a TDCR array and Defined Solid Angle Counting taking into account activity inhomogeneity in the active deposit. The results show consistency between the three methods within a limit of a 0.3%. An ampoule of this solution will be sent to the International Reference System (SIR) during 2009. Uncertainties were analysed and compared in detail for the three applied methods.